Command Quick Reference
Some [optional] commands have been included for clarity. All settings commands have a
corresponding query. See the
for SCPI.
:ACQuire (@chanlist)
Cancels any triggered measurements.
:DLOG
Stops internal data logging.
:ELOG (@chanlist)
Stops external data logging.
:HISTogram (@chanlist)
Stops the histogram current measurement. (only on N6781A/82A/85A/86A)
:TRANsient (@chanlist)
Cancels any triggered actions.
:COUNt?
Returns the number of times the unit has been calibrated.
:CURRent
[:LEVel] <
value
>, (@channel)
Calibrates the current programming.
:LIMit
:NEGative <
value
>, (@channel)
Calibrates negative current limit. (only on N678xA SMU, N6783A)
:POSitive <
value
>, (@channel)
Calibrates positive current limit. (only on N678xA SMU)
:MEASure <
value
>, (@channel)
Calibrates the current measurement.
:PEAK (@channel)
Calibrates peak current limit (only on N675xA, N676xA)
:DATA <
value
>
Enters the calibration value read by the external meter.
:DATE <
"date"
>, (@channel)
Stores the calibration date in nonvolatile memory.
:DPRog (@channel)
Calibrates the current downprogrammer.
:LEVel P1|P2|P3
Advances to the next level in the calibration.
:PASSword <
value
>
Sets a numeric password to prevent unauthorized calibration.
:RESistance 20 | 6, (@channel)
Calibrates the output resistance. (only on N6781A, N6785A)
:SAVE
Saves the calibration constants in non-volatile memory.
:STATe 0|OFF|1|ON
Enables or disables calibration mode.
:VOLTage
[:LEVel] <
value
>, (@channel)
Calibrates the voltage programming.
:CMRR, (@channel)
Calibrates the voltage common mode rejection ratio. (N675xA, N676xA)
:LIMit
:POSitive <
value
>, (@channel)
Calibrates positive current limit. (only on N678xA SMU)
:MEASure <
value
>, (@channel)
Calibrates the voltage measurement.
:AUXiliary, (@channel)
Calibrates the auxiliary voltage measurement. (only on N6781A, N6785A)
[:WINDow]
[:STATe] 0|OFF|1|ON
Turns the front panel display on or off.
:VIEW METER1 | METER4
Selects 1-channel or 4-channel meter view.
1 Quick Reference
26
Keysight N6705C Operating and Service Guide
Содержание N6705C
Страница 76: ......
Страница 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Страница 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Страница 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Страница 434: ......
Страница 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Страница 459: ...Overall Troubleshooting Keysight N6705C Operating and Service Guide 459 9 Service and Maintenance ...
Страница 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Страница 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Страница 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Страница 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Страница 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Страница 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Страница 476: ......