Marker View
Symbol/Field
Description
1 m1/m2 points
Shows where the measurement markers intersect the selected waveform. Data values at the bot-
tom of the display are referenced to the intersect locations of the markers. Calculations are based
on the data points in between the intersect locations.
2 Delta
Indicates the delta or absolute difference between the markers in units (volts, amps, or watts) and
in time (seconds). The value in parenthesis is the frequency, which is the reciprocal of the time
(1/time).
3 m2
Indicates the m2 marker value in volts, amps, or watts at the intersection point. Also indicates the
distance in time that the m2 marker is in relation to the present trigger position.
4 m1
Indicates the m1 marker value in volts, amps, or watts at the intersection point. Also indicates the
distance in time of the m1 marker in relation to the present trigger position.
5 Min
Indicates the minimum data value (in volts, amps, or watts) between the marker locations of the
selected waveform. Also indicates the distance in time of the minimum value in relation to the
present trigger position.
6 Avg
Calculates the average data value (in volts, amps, or watts) between the marker locations of the
selected waveform. Time indicates the time between markers over which the average value is
calculated.
7 V p-p
Indicates the maximum data value (in volts, amps, or watts) between the marker locations of the
selected waveform. Also indicates the distance in time of the maximum value in relation to the
present trigger position.
8 Trigger Mode
Calculates the difference between the maximum and minimum values. Time information is not
valid for calculated p-p values.
RMS (if selected)
Calculates the rms value between the marker locations. To view the rms values, you may need to
unselect one of the other measurements in the Scope Marker Properties window. Only 5
measurements may be displayed at a time.
4 Using the Measurement Functions
138
Keysight N6705C Operating and Service Guide
Содержание N6705C
Страница 76: ......
Страница 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Страница 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Страница 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Страница 434: ......
Страница 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Страница 459: ...Overall Troubleshooting Keysight N6705C Operating and Service Guide 459 9 Service and Maintenance ...
Страница 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Страница 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Страница 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Страница 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Страница 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Страница 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Страница 476: ......