SENSe:HISTogram:CURRent[:DC]:RANGe[:UPPer] <
value
>, (@<chanlist>)
SENSe:HISTogram:CURRent[:DC]:RANGe[:UPPer]? [MIN|MAX,] (@<chanlist>)
Selects the current histogram measurement range. The value that you enter must be the highest
value that you expect to measure. Units are in amperes. The instrument selects the range with the
best resolution for the value entered.
Parameter
Typical Return
0 to maximum|MIN|MAX, *RST MAX
<maximum value of range>
Selects the 1 A range: SENS:HIST:CURR:RANG 1, (@1)
l
When queried, the returned value is the maximum current that can be measured on the range that
is presently set.
SENSe:HISTogram:CURRent[:DC]:RANGe:AUTO 0|OFF|1|ON, (@<chanlist>)
SENSe:HISTogram:CURRent[:DC]:RANGe:AUTO? (@<chanlist>)
Enables or disables seamless histogram current measurement autoranging.
Parameter
Typical Return
0|OFF|1|ON, *RST OFF
0 or 1
Enable seamless histogram measurement autoranging:
SENS:HIST:CURR:RANG:AUTO ON, (@1)
l
Autoranging does not include the 10 μA range, which must be selected individually.
SENSe:HISTogram:FUNCtion:CURRent <Bool>, (@<chanlist>)
SENSe:HISTogram:FUNCtion:CURRent? (@<chanlist>)
Enables or disables the current histogram function.
Parameter
Typical Return
0|OFF|1|ON, *RST OFF
0 or 1
Enables current histogram measurements: SENS:HIST:FUNC:CURR ON, (@1)
SENSe:SWEep:OFFSet:POINts <
value
>, (@<chanlist>)
SENSe:SWEep:OFFSet:POINts? [MIN|MAX,] (@<chanlist>)
Defines the offset in a data sweep for triggered measurements. Positive values represent the delay
after the trigger occurs but before the samples are acquired. Negative values represent data samples
taken prior to the trigger.
7 SCPI Programming Reference
324
Keysight N6705C Operating and Service Guide
Содержание N6705C
Страница 76: ......
Страница 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Страница 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Страница 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Страница 434: ......
Страница 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Страница 459: ...Overall Troubleshooting Keysight N6705C Operating and Service Guide 459 9 Service and Maintenance ...
Страница 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Страница 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Страница 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Страница 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Страница 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Страница 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Страница 476: ......