It takes a few milliseconds for the instrument to be ready to receive a trigger signal after receiving the
INITiate:TRANsient command. If a trigger occurs before the trigger system is ready for it, the trigger
will be ignored. You can test the WTG_tran bit in the operation status register to know when the
instrument is ready to receive a trigger after being initiated.
To query the WTG_tran bit (bit 4):
If a bit value of 16 is returned in the query, the WTG_tran bit is true, and the instrument is ready to
receive the trigger signal. Refer to the N6705 Programmer’s Reference Help file for more information.
Unless INITiate:CONTinuous:TRANsient is programmed, it will be necessary to initiate
the transient trigger system each time a triggered action is desired.
To trigger the Arb if the trigger source is BUS:
As previously discussed, a trigger can also be generated by a trigger signal applied to the rear panel
trigger input BNC connector. If this is configured as the trigger source, the instrument will wait
indefinitely for the trigger signal. If the trigger does not occur, you must manually return the trigger
system to the idle state.
If the trigger does not occur, and you wish to cancel the arbitrary waveform, send
When a trigger is received, the Arb runs. When the Arb completes, the trigger system returns to the
idle state. You can test the TRAN_active bit in the operation status register to know when the transient
trigger system has returned to the idle state.
To query the TRAN_active bit (bit 6):
If a bit value of 64 is returned in the query, the TRAN_active bit is true, and the Arb has NOT completed.
When the TRAN_active bit is false, the Arb has completed. Refer to the N6705 Programmer’s
Reference Help file for more information.
Importing/Exporting Arb Data
You can only import and export User-Defined and Constant-Dwell Arb Data.
You can create an arbitrary waveform in a Microsoft Excel spreadsheet and import it into the
instrument using the Import function as previously explained in this section under
and
Configuring Constant-Dwell Properties
. Likewise, you can also Export an
arbitrary waveform from the instrument to a spreadsheet.
Keysight N6705C Operating and Service Guide
107
3 Using the Source Functions
Содержание N6705C
Страница 76: ......
Страница 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Страница 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Страница 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Страница 434: ......
Страница 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Страница 459: ...Overall Troubleshooting Keysight N6705C Operating and Service Guide 459 9 Service and Maintenance ...
Страница 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Страница 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Страница 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Страница 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Страница 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Страница 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Страница 476: ......