Low Range Voltage Measurement - Models N6781A, N6782A, N6784A
Step 1. Select the low-voltage measurement range. The value entered must be the maximum voltage
of the range to be calibrated. This example selects the 1 V range of output 1.
Front Panel Menu Reference
SCPI Command
From the Calibration menu, select Voltage, Low Range Measurement. Then select
Start.
Step 2. Select the first voltage calibration point and enter the data. This value may be positive or
negative.
Front Panel Menu Reference
SCPI Command
Display shows: “Enter P1 measured data”Enter the data from the external DMM.
Press Select when done.
CAL:LEV P1
*OPC?
CAL:DATA <data>
Step 3. Select the second voltage calibration point and enter the data.
Front Panel Menu Reference
SCPI Command
Display shows: “Enter P2 measured data”Enter the data from the external
DMM.Press Select when done.
Press Next to finish calibration.
CAL:LEV P2
*OPC?
CAL:DATA <data>
Step 4 applies only to models N6781A and N6782A
Step 4. Repeat steps 1 through 3 for the 100 mV range. This time, select the Low2 range.
Steps 5 and 6 apply only to model N6784A
Step 5. Select the third voltage calibration point for the negative range and enter the data. This value
should be negative.
Front Panel Menu Reference
SCPI Command
Display shows: “Enter P3 measured data”
Enter the data from the external DMM. Press Select when done. Press Next to
finish calibration.
CAL:LEV P3
*OPC?
CAL:DATA <data>
Step 6. Repeat steps 1 through 3, and 5 for the 100 mV range. This time, select the Low2 range.
Positive Voltage Limit - Models N6781A, N6782A, N6784A
Step 1. Select the positive voltage limit range. The value to program a range must be the maximum
voltage of the range to be calibrated. This example selects the 20 V positive limit range of output 1.
Front Panel Menu Reference
SCPI Command
From the Calibration menu, select Voltage, Positive Voltage Limit. Then select
Start. Select the 20 V range.
Keysight N6705C Operating and Service Guide
401
8 Verification and Calibration
Содержание N6705C
Страница 76: ......
Страница 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Страница 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Страница 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Страница 434: ......
Страница 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Страница 459: ...Overall Troubleshooting Keysight N6705C Operating and Service Guide 459 9 Service and Maintenance ...
Страница 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Страница 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Страница 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Страница 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Страница 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Страница 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Страница 476: ......