Execution Errors (these errors set Standard Event Status register bit #4)
-200 Execution error
Generic syntax error
-220 Parameter error
A data element related error occurred.
-221 Settings conflict
A data element could not be executed because of the present instrument state.
-222 Data out of range
A data element could not be executed because the value was outside the valid range.
-223 Too much data
A data element was received that contains more data than the instrument can handle.
-224 Illegal parameter value
An exact value was expected but not received.
-225 Out of memory
The device has insufficient memory to perform the requested operation.
−226 Lists not same length
One or more lists are not the same length.
-230 Data corrupt or stale
Possible invalid data. A new reading was started but not completed.
-231 Data questionable
The measurement accuracy is suspect.
-232 Invalid format
The data format or structure is inappropriate.
-233 Invalid version
The version of the data format is incorrect to the instrument.
-240 Hardware error
The command could not be executed because of a hardware problem with the instrument.
-241 Hardware missing
The command could not be executed because of missing hardware, such as an option.
−260 Expression error
Keysight N6705C Operating and Service Guide
367
7 SCPI Programming Reference
Содержание N6705C
Страница 76: ......
Страница 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Страница 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Страница 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Страница 434: ......
Страница 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Страница 459: ...Overall Troubleshooting Keysight N6705C Operating and Service Guide 459 9 Service and Maintenance ...
Страница 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Страница 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Страница 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Страница 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Страница 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Страница 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Страница 476: ......