SENSe:DLOG:TINTerval <
value
>
SENSe:DLOG:TINTerval? [MIN|MAX,]
Sets the integration time of the datalog measurement. Values from 1 millisecond to 60 seconds can
be specified
This command has been superseded by SENSe:DLOG:PERiod, but is still available for
backward compatibility.
Parameter
Typical Return
0.0002048 to 60|MIN|MAX, *RST 0.1
<datalog period>
Specifies a datalog time interval of 0.01 seconds: SENS:DLOG:TINT 0.01, (@1)
l
This function cannot be set to values less than 75 milliseconds when interleaved data logging is in
effect.
SENSe:ELOG:CURRent[:DC]:RANGe[:UPPer] <
value
>, (@<chanlist>)
SENSe:ELOG:CURRent[:DC]:RANGe[:UPPer]? [MIN|MAX,] (@<chanlist>)
SENSe:ELOG:VOLTage[:DC]:RANGe[:UPPer] <
value
>, (@<chanlist>)
SENSe:ELOG:VOLTage[:DC]:RANGe[:UPPer]? [MIN|MAX,] (@<chanlist>)
Selects the external datalog (Elog) current or voltage measurement range. The value that you enter
must be the highest value that you expect to measure. Units are in amperes or volts. The instrument
selects the range with the best resolution for the value entered.
Parameter
Typical Return
0 to maximum|MIN|MAX, *RST MAX
<maximum value of range>
Selects the 1 A range: SENS:ELOG:CURR:RANG 1, (@1)
Selects the 2 V range: SENS:ELOG:VOLT:RANG 2, (@1)
l
When queried, the returned value is the maximum current or voltage that can be measured on the
range that is presently set.
SENSe:ELOG:CURRent[:DC]:RANGe:AUTO 0|OFF|1|ON, (@<chanlist>)
SENSe:ELOG:CURRent[:DC]:RANGe:AUTO? (@<chanlist>)
SENSe:ELOG:VOLTage[:DC]:RANGe:AUTO 0|OFF|1|ON , (@<chanlist>)
SENSe:ELOG:VOLTage[:DC]:RANGe:AUTO? (@<chanlist>)
Enables/disables seamless Elog measurement autoranging.
7 SCPI Programming Reference
320
Keysight N6705C Operating and Service Guide
Содержание N6705C
Страница 76: ......
Страница 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Страница 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Страница 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Страница 434: ......
Страница 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Страница 459: ...Overall Troubleshooting Keysight N6705C Operating and Service Guide 459 9 Service and Maintenance ...
Страница 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Страница 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Страница 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Страница 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Страница 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Страница 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Страница 476: ......