[SOURce:]ARB:SEQuence:STEP:CURRent:TRAPezoid:FTIMe <
value
>, <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:TRAPezoid:FTIMe? [MIN|MAX,] <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:TRAPezoid:FTIMe <
value
>, <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:TRAPezoid:FTIMe? [MIN|MAX,] <
step#
>, (@<chanlist>)
Specifies the fall time of the trapezoid in seconds for Arb sequence steps.
Parameter
Typical Return
0 – 262.144 | MIN | MAX, *RST 1
<fall time>
Step 0 - 99, *RST 0
(none)
Programs a fall time for the current trapezoid: ARB:CURR:TRAP:FTIM 10,(@1)
Programs a fall time for the voltage trapezoid: ARB:VOLT:TRAP:FTIM 10,(@1)
Programs a fall time for step 0:ARB:SEQ:STEP:CURR:TRAP:FTIM 10, 0,(@1)
[SOURce:]ARB:CURRent:TRAPezoid:RTIMe <
value
>, (@<chanlist>)
[SOURce:]ARB:CURRent:TRAPezoid:RTIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:TRAPezoid:RTIMe<
value
>, (@<chanlist>)
[SOURce:]ARB:VOLTage:TRAPezoid:RTIMe? [MIN|MAX,] (@<chanlist>)
Specifies the rise time of the trapezoid in seconds. Referenced to t1 in the Trapezoid diagrams.
[SOURce:]ARB:SEQuence:STEP:CURRent:TRAPezoid:RTIMe <
value
>, <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:TRAPezoid:RTIMe? [MIN|MAX,] <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:TRAPezoid:RTIMe <
value
>, <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:TRAPezoid:RTIMe? [MIN|MAX,] <
step#
>, (@<chanlist>)
Specifies the rise time of the trapezoid in seconds for Arb sequence steps.
Parameter
Typical Return
0 – 262.144 | MIN | MAX, *RST 1
<rise time>
Step 0 - 99, *RST 0
(none)
Programs a rise time for the current trapezoid: ARB:CURR:TRAP:RTIM 10,(@1)
Programs a rise time for the voltage trapezoid: ARB:VOLT:TRAP:RTIM 10,(@1)
Programs a rise time for step 0:ARB:SEQ:STEP:CURR:TRAP:RTIM 10, 0,(@1)
[SOURce:]ARB:CURRent:TRAPezoid:STARt[:LEVel] <
value
>, (@<chanlist>)
[SOURce:]ARB:CURRent:TRAPezoid:STARt[:LEVel]? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:TRAPezoid:STARt[:LEVel] <
value
>, (@<chanlist>)
[SOURce:]ARB:VOLTage:TRAPezoid:STARt[:LEVel]? [MIN|MAX,] (@<chanlist>)
Specifies the initial and final current or voltage level of the trapezoid. Referenced to I0 and V0 in the
Trapezoid diagrams.
7 SCPI Programming Reference
256
Keysight N6705C Operating and Service Guide
Содержание N6705C
Страница 76: ......
Страница 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Страница 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Страница 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Страница 434: ......
Страница 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Страница 459: ...Overall Troubleshooting Keysight N6705C Operating and Service Guide 459 9 Service and Maintenance ...
Страница 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Страница 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Страница 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Страница 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Страница 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Страница 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Страница 476: ......