[SOURce:]VOLTage:LIMit[:POSitive][:IMMediate][:AMPLitude] <
value
> , (@<chanlist>)
[SOURce:]VOLTage:LIMit[:POSitive][:IMMediate][:AMPLitude]? [MIN|MAX,] (@<chanlist>)
[SOURce:]VOLTage:LIMit:NEGative[:IMMediate][:AMPLitude] <
value
>, (@<chanlist>)
[SOURce:]VOLTage:LIMit:NEGative[:IMMediate][:AMPLitude]? [MIN|MAX,] (@<chanlist>)
Sets the voltage limit when in current priority mode. Units are in volts. The maximum value depends
on the voltage rating of the unit. The minimum value is the most negative value.
Parameter
Typical Return
Positive: 0 to 102% of rating|MIN|MAX, *RST 102%
Negative: -102% of rating to 0|MIN|MAX, *RST -102%
<+ voltage limit>
<- voltage limit>
Sets the positive voltage limit to 20.4 V: VOLT:LIM 20, (@1)
Sets the negative voltage limit to -20.4 V: VOLT:LIM:NEG -20, (@1)
l
MIN refers to the most negative value of the range. MAX refers to the least negative value of the
range.
l
The query returns the programmed voltage limit in the form +n.nn for each channel spe-
cified. Multiple responses are separated by commas.
[SOURce:]VOLTage:LIMit:COUPle 0|OFF|1|ON, (@<chanlist>)
[SOURce:]VOLTage:LIMit:COUPle? (@<chanlist>)
Sets the voltage limit tracking state. The negative voltage limit tracks the positive voltage limit
setting.
Parameter
Typical Return
0|OFF|1|ON, *RST OFF
0 or 1
Enables the voltage limit tracking state: VOLT:LIM:COUP ON, (@1)
[SOURce:]VOLTage:MODE FIXed|STEP|LIST|ARB, (@<chanlist>)
[SOURce:]VOLTage:MODE? (@<chanlist>)
Sets the transient mode. This determines what happens to the output voltage when the transient
system is initiated and triggered.
FIXed - keeps the output voltage at its immediate value.
STEP - steps the output to the triggered level when a trigger occurs.
LIST - causes the output to follow the list values when a trigger occurs.
ARB - causes the output to follow the arbitrary waveform values when a trigger occurs.
7 SCPI Programming Reference
344
Keysight N6705C Operating and Service Guide
Содержание N6705C
Страница 76: ......
Страница 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Страница 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Страница 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Страница 434: ......
Страница 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Страница 459: ...Overall Troubleshooting Keysight N6705C Operating and Service Guide 459 9 Service and Maintenance ...
Страница 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Страница 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Страница 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Страница 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Страница 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Страница 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Страница 476: ......