Configuring User-Defined Arbs
User-defined Arbs can contain up to 511 voltage or current steps that are entered individually in the
User-defined Properties window.
You can also populate the User-defined voltage or current Arb with values from a previously
configured “standard” arbitrary waveform and then edit the steps in the User-defined Properties
window.
To convert one of the standard Arbs, program the standard Arb parameters and then select the Edit
Points button to populate the User-defined Arb with the values from the standard Arb.
You can also import a User-defined Arb that was previously created using a spreadsheet as explained
later in this section under
Importing and Exporting Arb Data
.
Step 1 – Select a Voltage or Current User-defined Arb:
Press the Arb key twice or press the Arb key and then Properties to access the Arb Selection window.
In the Arb Type dropdown menu, select either a Voltage or Current Arb type. Then select the Pulse
output type.
Step 2 – Configure the User-defined properties
Press the Properties key or select the Arb Properties button.
For step 0, enter the voltage or current values. Then enter the time or dwell for the step. Check the
Trigger box if you wish to generate an external trigger signal at the start of the step.
Select the Add button to insert a new step below the selected step. Note that the values in the new
step are copied from the previous step. Edit the values in the step. Select Clear if you want to clear all
values. Select Delete if you want to delete the selected step. Continue to Add steps until your
waveform is complete. Use the
▲ ▼
navigation keys to move through the list.
Specify what happens when the waveform completes – select whether the output returns to the DC
value that was in effect before the waveform started, or whether the output should remain at the last
Arb value.
Keysight N6705C Operating and Service Guide
95
3 Using the Source Functions
Содержание N6705C
Страница 76: ......
Страница 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Страница 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Страница 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Страница 434: ......
Страница 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Страница 459: ...Overall Troubleshooting Keysight N6705C Operating and Service Guide 459 9 Service and Maintenance ...
Страница 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Страница 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Страница 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Страница 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Страница 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Страница 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Страница 476: ......