Chapter 10
195
Command Reference
E4981A SCPI Commands
10
. Comm
and
Refe
ren
ce
:DATA:FEED:BUF2
Syntax
:DATA:FEED:BUF2 {“CALCulate1”|”CALCulate2”|””}
:DATA:FEED:BUF2?
Description
Selects the measurement data fed into data buffer 2from the following: primary parameter,
secondary parameter, or none to be fed. The query response is a string with double
quotation marks (“).
NOTE
The functionality of this command is same as
with value of
BUF2.
Parameters
Query response
{"CALCulate1"|"CALCulate2"|""}<newline><^END>
Related commands
:DATA:FEED:CONTrol[:STATe] on page 198
:DATA:POINts:BUF1 on page 200
Equivalent key
sequence
No equivalent key is available on the front panel.
Description
“CALCulate1”
Specifies the primary parameter as the measurement data fed into
the data buffer 2.
“CALCulate2”
Specifies the secondary parameter as the measurement data fed
into the data buffer 2.
““ (initial value)
Does not feed the measurement data into the data buffer 2.
Содержание E4981A
Страница 2: ......
Страница 20: ...18 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 60: ...58 Chapter4 Preparation for Accurate Measurement Executing Correction Using the Cable Correction Function ...
Страница 88: ...86 Chapter7 Sorting Based on Measured Result Comparator Function Sample Program ...
Страница 162: ...160 Chapter9 Measurement Applications Sample Programs Detecting Error Occurrence using SRQ HT Basic ...
Страница 376: ...374 AppendixE Error Messages Warning Messages WARNING ...
Страница 382: ......
Страница 383: ...This information is subject to change without notice Keysight Technologies 2008 2014 Edition 2 August 2014 E4981 90001 ...