Appendix E
369
Error Messages
Error messages (order of error number)
E.
Er
ro
r Me
ssag
es
-141
Invalid character data
A character data element has been recieved where E4981A does not accepts any character
data element.
-148
Character data not allowed
Character data not allowed for this operation.
-150
String data error
An error not included in the error numbers between -151 and -159 has occurred during the
syntax analysis of a string data element.
-151
Invalid string data
Character string data are expected, but the string data received are invalid for some reason.
(Refer to IEEE488.2,7.7.5.2.) For example, the END message is received before the end
quotation mark character appears.
-158
String data not allowed
A string data element has been received where the E4981A does not accept any string data
element. For example, a parameter must be enclosed with double quotation marks (“...”)
but they are missing.
-161
Invalid block data
Block data are expected, but the block data received are invalid for some reason. (Refer to
IEEE488.2,7.7.6.2.) For example, the END message is received before the length of the
block data is reached.
-168
Block data not allowed
A block data element has been received where the E4981A does not accept any block data
element.
-170
Expression error
An error not included in the error numbers between -171 and -179 has occurred during the
syntax analysis of equation data.
-171
Invalid expression
The equation data element is invalid. (Refer to IEEE488.2,7.7.7.2.) For example,
parentheses are not paired or a character violates the standard.
-178
Expression data not allowed
An equation data element has been received where the E4981A does not accept any
equation data element.
-200
Execution error
A comprehensive execution error has occurred for which the E4981A cannot detect further
details. This error code simply indicates the occurrence of an execution error that is defined
in IEEE488.2,11.5.1.1.5.
-211
Trigger ignored
A trigger command or trigger signal has been received and recognized by the E4981A, but
it is ignored due to the timing relationship with the E4981A. For example, this happens
when the E4981A’s trigger system is not in the Waiting for Trigger state.
Содержание E4981A
Страница 2: ......
Страница 20: ...18 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 60: ...58 Chapter4 Preparation for Accurate Measurement Executing Correction Using the Cable Correction Function ...
Страница 88: ...86 Chapter7 Sorting Based on Measured Result Comparator Function Sample Program ...
Страница 162: ...160 Chapter9 Measurement Applications Sample Programs Detecting Error Occurrence using SRQ HT Basic ...
Страница 376: ...374 AppendixE Error Messages Warning Messages WARNING ...
Страница 382: ......
Страница 383: ...This information is subject to change without notice Keysight Technologies 2008 2014 Edition 2 August 2014 E4981 90001 ...