298
Chapter 10
Command Reference
E4981A SCPI Commands
:TRIGger[:SEQ1][:IMMediate]
Syntax
:TRIGger[:SEQ1][:IMMediate]
Description
Immediately generates a trigger and executes a measurement. This command is valid for
trigger source values of MANual or BUS trigger specified by
:TRIGger[:SEQ1]:SOURce
command. On setting the trigger source value to INTernal or EXTernal, this command
generates an error.
If the trigger system is not in the trigger wait state (trigger event detection state), executing
this command causes an error and the command is ignored.
For details on the trigger system, refer to
. (No query)
Equivalent key
sequence
[
Trigger
]
:TRIGger[:SEQ1]:SLOPe
Syntax
:TRIGger[:SEQ1]:SLOPe {POSitive|NEGative}
:TRIGger[:SEQ1]:SLOPe?
Description
Determines the trigger polarity for BNC external trigger on the rear panel. This does not
affects the trigger signal on handler and scanner interface.
Related commands
Parameters
NOTE
The value of this command is not changed with
and is only
changed with Factory Default reset (available only through Front Panel).
Query response
{POS|NEG}<newline><^END>
Equivalent key
sequence
[
System
] - EXT TRIG POL - POS|NEG
Description
POSitive (Initial
[Factory set] value)
Detects positive edge trigger signal.
NEGative
Detects negative edge trigger signal.
Содержание E4981A
Страница 2: ......
Страница 20: ...18 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 60: ...58 Chapter4 Preparation for Accurate Measurement Executing Correction Using the Cable Correction Function ...
Страница 88: ...86 Chapter7 Sorting Based on Measured Result Comparator Function Sample Program ...
Страница 162: ...160 Chapter9 Measurement Applications Sample Programs Detecting Error Occurrence using SRQ HT Basic ...
Страница 376: ...374 AppendixE Error Messages Warning Messages WARNING ...
Страница 382: ......
Страница 383: ...This information is subject to change without notice Keysight Technologies 2008 2014 Edition 2 August 2014 E4981 90001 ...