Chapter 9
105
Measurement Applications (Sample Programs)
Executing OPEN/SHORT/LOAD Correction
9.
M
easu
rem
ent
Applica
tio
ns
(S
amp
le
Prog
ra
ms)
Result = ""
' Open Correction Values
ErrorCheck viVPrintf(Agte4981a, ":CORR:DATA? STAN1" + vbLf, 0)
ErrorCheck viVScanf(Agte4981a, "%t", Result)
Res = Split(Result, ",")
Open1 = Val(Res(0))
Open2 = Val(Res(1))
Result = ""
' Short Correction Parameter
ErrorCheck viVPrintf(Agte4981a, ":CORR:CKIT:STAN2:FORMAT?" +
vbLf, 0) ' Read Primary and Secondary Load Types CP and D
ErrorCheck viVScanf(Agte4981a, "%t", Result)
StdShort = Mid(Result, 1, 2)
Result = ""
' Short Correction Values
ErrorCheck viVPrintf(Agte4981a, ":CORR:DATA? STAN2" + vbLf, 0)
ErrorCheck viVScanf(Agte4981a, "%t", Result)
Res = Split(Result, ",")
Short1 = Val(Res(0))
Short2 = Val(Res(1))
Result = ""
' Load Correction Parameter
ErrorCheck viVPrintf(Agte4981a, ":CORR:CKIT:STAN3:FORMAT?" +
vbLf, 0) ' Read Primary and Secondary Load Types CP and D
ErrorCheck viVScanf(Agte4981a, "%t", Result)
StdLoad = Mid(Result, 1, 3)
Result = ""
Содержание E4981A
Страница 2: ......
Страница 20: ...18 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 60: ...58 Chapter4 Preparation for Accurate Measurement Executing Correction Using the Cable Correction Function ...
Страница 88: ...86 Chapter7 Sorting Based on Measured Result Comparator Function Sample Program ...
Страница 162: ...160 Chapter9 Measurement Applications Sample Programs Detecting Error Occurrence using SRQ HT Basic ...
Страница 376: ...374 AppendixE Error Messages Warning Messages WARNING ...
Страница 382: ......
Страница 383: ...This information is subject to change without notice Keysight Technologies 2008 2014 Edition 2 August 2014 E4981 90001 ...