Chapter 10
295
Command Reference
E4981A SCPI Commands
10
. Comm
and
Refe
ren
ce
:TEST:SCANner:MODE
Syntax
:TEST:SCANner:MODE {ON|OFF|1|0}
:TEST:SCANner:MODE?
Description
This command sets the Scanner Test Mode ON/OFF for test purpose and can be used to
check/troubleshoot the Test pin signal of Scanner interface.
NOTE
is set to ON, the interface pin signal value can be
controlled/read by :TEST:SCAN:xxxx commands. After completing the test,
should set to OFF so that scanner interface pin signal is generated
according to the actual measurement result.
Parameters
Query response
{1|0}<newline><^END>
Related commands
:TEST:SCANner:CH? on page 292
:TEST:SCANner:EOM on page 293
:TEST:SCANner:INDex on page 294
:TEST:SCANner:TRIGger? on page 296
:TEST:SCANner:VALid? on page 296
Equivalent key
sequence
No equivalent key is available on the front panel.
Description
ON
Sets Scanner mode to ON
OFF
Sets Scanner mode to OFF.
Содержание E4981A
Страница 2: ......
Страница 20: ...18 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 60: ...58 Chapter4 Preparation for Accurate Measurement Executing Correction Using the Cable Correction Function ...
Страница 88: ...86 Chapter7 Sorting Based on Measured Result Comparator Function Sample Program ...
Страница 162: ...160 Chapter9 Measurement Applications Sample Programs Detecting Error Occurrence using SRQ HT Basic ...
Страница 376: ...374 AppendixE Error Messages Warning Messages WARNING ...
Страница 382: ......
Страница 383: ...This information is subject to change without notice Keysight Technologies 2008 2014 Edition 2 August 2014 E4981 90001 ...