Chapter 10
283
Command Reference
E4981A SCPI Commands
10
. Comm
and
Refe
ren
ce
:TEST:HANDler:BIN
Syntax
:TEST:HANDler:BIN <numeric>
Description
This command sets the Handler Bin No for test purpose and can be used to
check/troubleshoot the Test pin signal of Handler interface. (No query)
NOTE
is set to ON, the interface pin signal value can be
controlled/read by: TEST:HAND:xxxx commands. After completing the test,
should set to OFF so that handler interface pin signal is generated
according to the actual measurement result.
Parameters
Related Commands
:TEST:HANDler:COMP on page 284
:TEST:HANDler:KEYLock? on page 284
:TEST:HANDler:MODE on page 285
Equivalent key
sequence
No equivalent key is available on the front panel.
Numeric Value
Description
0
Out of Bins
1-9
BIN 1 to 9
10
AUX BIN
11
All Off
Содержание E4981A