256
Chapter 10
Command Reference
E4981A SCPI Commands
[:SENSe][:FIMPedance]:CONTact1:VERify[:STATe]
Syntax
[:SENSe][:FIMPedance]:CONTact1:VERify[:STATE] {ON|OFF|1|0}
[:SENSe][:FIMPedance]:CONTact1:VERify[:STATe]?
Description
Determines whether to use the contact check function.
NOTE
Contact check function is not available at 1 MHz measurement frequency.
Parameters
Query response
{1|0}<newline><^END>
Related commands
[:SENSe][:FIMPedance]:CONTact1:VERify:THReshold1 on page 257
[:SENSe][:FIMPedance]:CONTact1:VERify:THReshold2 on page 258
Equivalent key
sequence
[
Meas Setup
] - CONT CHK1- ON|OFF
Description
ON or 1
Enables the contact check function.
OFF or 0 (initial value) Disables the contact check function.
Содержание E4981A
Страница 2: ......
Страница 20: ...18 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 60: ...58 Chapter4 Preparation for Accurate Measurement Executing Correction Using the Cable Correction Function ...
Страница 88: ...86 Chapter7 Sorting Based on Measured Result Comparator Function Sample Program ...
Страница 162: ...160 Chapter9 Measurement Applications Sample Programs Detecting Error Occurrence using SRQ HT Basic ...
Страница 376: ...374 AppendixE Error Messages Warning Messages WARNING ...
Страница 382: ......
Страница 383: ...This information is subject to change without notice Keysight Technologies 2008 2014 Edition 2 August 2014 E4981 90001 ...