80
Chapter 7
Sorting Based on Measured Result (Comparator Function)
Setting Up Comparator Function
Turning ON/OFF AUX BIN function
To turn ON/OFF the AUX BIN function, use the following command.
•
:CALCulate1:COMParator:AUXBin on page 175
Depending on the ON/OFF state of the AUX BIN function, the sorting result when the
value of the secondary parameter exceeds the limit range varies as shown in
.
Setting up condition to make a beep sound
You can select the condition that produces a beep sound based on the comparator sorting
judgment result from the following:
• Beep when the sorting judgment result is OUT_OF_BINS or AUX_BIN
• Beep when the sorting judgment result is BIN1 to BIN9
To set the condition that makes a beep sound, use the following command.
•
:CALCulate1:COMParator:BEEPer:CONDition on page 175
You can also disable the beep sound. To turn ON/OFF the beep, use either of the following
commands (they have the same function).
•
:CALCulate1:COMParator:BEEPer[:STATe] on page 176
•
:SYSTem:BEEPer:STATe on page 269
Table 7-1
Sorting result when measured secondary parameter value exceeds limit range
Primary parameter sorting result
AUX BIN function
Sorting result
One of BIN1 to BIN9
OFF
OUT_OF_BINS
ON
AUX_BIN
Not sorted to any BIN
Not applicable
OUT_OF_BINS
Содержание E4981A
Страница 2: ......
Страница 20: ...18 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 60: ...58 Chapter4 Preparation for Accurate Measurement Executing Correction Using the Cable Correction Function ...
Страница 88: ...86 Chapter7 Sorting Based on Measured Result Comparator Function Sample Program ...
Страница 162: ...160 Chapter9 Measurement Applications Sample Programs Detecting Error Occurrence using SRQ HT Basic ...
Страница 376: ...374 AppendixE Error Messages Warning Messages WARNING ...
Страница 382: ......
Страница 383: ...This information is subject to change without notice Keysight Technologies 2008 2014 Edition 2 August 2014 E4981 90001 ...