166
Chapter 10
Command Reference
IEEE Common Commands
*OPC?
Syntax
*OPC?
Description
1 is read out at the completion of all pending operations. (Query only)
Query response
{1}<newline><^END>
Equivalent key
sequence
No equivalent key is available on the front panel.
*OPT?
Syntax
*OPT?
Description
Reads out the identification number of an option installed in the E4981A. (Query only)
Query response
{numeric}<newline><^END>
NOTE
Option 001 or 002 is available for E4981A.
Equivalent key
sequence
[System]
*RCL
Syntax
*RCL <numeric>
Description
Recalls the instrument setups saved in the register of the specified number on Flash
(internal) memory or USB pen drive (external memory). For details of recalled instrument
setups, refer to
Operation Manual, Appendix C “Initial Setting”
.(No query)
Parameters
If the specified parameter is out of the allowable setup range, an error occurs.
Related commands
Equivalent key
sequence
[
Save/Recall
] - No {1-9} - RECALL
<numeric>
Description
Specified number
Range
0 to 9 (Internal Flash Memory)
10 to 19 (External USB Memory)
Resolution
1
Содержание E4981A
Страница 2: ......
Страница 20: ...18 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 60: ...58 Chapter4 Preparation for Accurate Measurement Executing Correction Using the Cable Correction Function ...
Страница 88: ...86 Chapter7 Sorting Based on Measured Result Comparator Function Sample Program ...
Страница 162: ...160 Chapter9 Measurement Applications Sample Programs Detecting Error Occurrence using SRQ HT Basic ...
Страница 376: ...374 AppendixE Error Messages Warning Messages WARNING ...
Страница 382: ......
Страница 383: ...This information is subject to change without notice Keysight Technologies 2008 2014 Edition 2 August 2014 E4981 90001 ...