:CALCulate1:MATH:STATe
Syntax
:CALCulate1:MATH:STATe {ON|OFF|1|0}
:CALCulate1:MATH:STATe?
Description
Determines whether to use the function (deviation measurement mode) that displays the
primary parameter measurement result in deviation from the reference value (set using the
NOTE
Changing any of the measurement parameters will automatically disable this function. To
avoid this, you have to write a program so that this command is executed after the
execution of the measurement parameter setup command (
Parameters
Query response
{1|0}<newline><^END>
Related commands
:CALCulate1:MATH:EXPRession:NAME on page 189
:CALCulate1:FORMat on page 188
:CALCulate2:FORMat on page 191
:CALCulate2:MATH:STATe on page 193
Equivalent key
sequence
[
Meas Setup
] - DEV A - OFF
Description
ON or 1
Enables the deviation measurement mode.
OFF or 0 (initial value)
Disables the deviation measurement mode (in other
words, displays the measurement result in an absolute
value).
Содержание E4981A
Страница 2: ......
Страница 20: ...18 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 60: ...58 Chapter4 Preparation for Accurate Measurement Executing Correction Using the Cable Correction Function ...
Страница 88: ...86 Chapter7 Sorting Based on Measured Result Comparator Function Sample Program ...
Страница 162: ...160 Chapter9 Measurement Applications Sample Programs Detecting Error Occurrence using SRQ HT Basic ...
Страница 376: ...374 AppendixE Error Messages Warning Messages WARNING ...
Страница 382: ......
Страница 383: ...This information is subject to change without notice Keysight Technologies 2008 2014 Edition 2 August 2014 E4981 90001 ...