Chapter 10
229
Command Reference
E4981A SCPI Commands
10
. Comm
and
Refe
ren
ce
[:SENSe]:AVERage[:STATe]
Syntax
[:SENSe]:AVERage[:STATe] {ON|OFF|1|0}
[:SENSe]:AVERage[:STATe]?
Description
Turns ON/OFF the averaging function.
Parameters
Query response
{1|0}<newline><^END>
Related commands
[:SENSe]:AVERage:COUNt on page 228
Equivalent key
sequence
[
Meas Setup
] - AVG - ON/OFF
NOTE
Setting the averaging count with the front panel keys will automatically turn on the
function.
[:SENSe]:CORRection:CKIT:STAN1:FORMat
Syntax
[:SENSe]:CORRection:CKIT:STAN1:FORMat {GB|CPG}
[:SENSe]:CORRection:CKIT:STAN1:FORMat?
Description
Sets the parameter types of the OPEN correction data.
Parameters
Query response
{GB|CPG}<newline><^END>
Related commands
[:SENSe]:CORRection:DATA on page 236
Equivalent key
sequence
[
Meas Setup
] - CORRECTION - OPEN - G-B/Cp-G
Description
ON or 1 (initial value)
Turns ON the averaging function.
OFF or 0
Turns OFF the averaging function.
Description
GB (initial value)
Specifies G as the primary parameter and B as the secondary
parameter.
CPG
Specifies Cp as the primary parameter and G as the secondary
parameter.
Содержание E4981A
Страница 2: ......
Страница 20: ...18 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 60: ...58 Chapter4 Preparation for Accurate Measurement Executing Correction Using the Cable Correction Function ...
Страница 88: ...86 Chapter7 Sorting Based on Measured Result Comparator Function Sample Program ...
Страница 162: ...160 Chapter9 Measurement Applications Sample Programs Detecting Error Occurrence using SRQ HT Basic ...
Страница 376: ...374 AppendixE Error Messages Warning Messages WARNING ...
Страница 382: ......
Страница 383: ...This information is subject to change without notice Keysight Technologies 2008 2014 Edition 2 August 2014 E4981 90001 ...