236
Chapter 10
Command Reference
E4981A SCPI Commands
[:SENSe]:CORRection:DATA
Syntax
[:SENSe]:CORRection:DATA {STAN1|STAN2|STAN3},<numeric 1>,<numeric 2>
[:SENSe]:CORRection:DATA? {STAN1|STAN2|STAN3}
Description
Sets the correction data for OPEN/SHORT/LOAD correction.
These correction data are measured for the measurement frequency used in executing the
command (set with the
command).
When using the multi-correction function (set to ON with the
[:SENSe]:CORRection:MULTiple[:STATe]
command), this is set as the correction data for
the selected channel when executing the command (selected with the
[:SENSe]:CORRection:MULTiple:CHANnel
To activate the set correction data, use the
[:SENSe]:CORRection:OPEN[:STATe]
[:SENSe]:CORRection:SHORt[:STATe]
[:SENSe]:CORRection:LOAD[:STATe]
commands to turn on the OPEN/SHORT/LOAD correction function.
The transfer format of data read out with this command conforms to the setup of the
:FORMat[:DATA]
command.
Parameters
• When specifying STAN1 as parameter 1:
• When specifying STAN2 as parameter 1:
Parameter 1: {STAN1|STAN2|STAN3}
STAN1
Sets or reads out the data for OPEN correction.
STAN2
Sets or reads out the data for SHORT correction.
STAN3
Sets or reads out the data for LOAD correction.
Parameter 2: <numeric 1>
Parameter 3: <numeric 2>
Description
Value of the primary parameter
*1
.
Value of the secondary parameter
*1
.
Range
-99.9999E9 to 99.9999E9
-99.9999E9 to 99.9999E9
Initial value
0
0
Unit
Depends on the type of parameter. Depends on the type of parameter.
*1.Use the
[:SENSe]:CORRection:CKIT:STAN1:FORMat
command to specify the type
of parameter.
Parameter 2: <numeric 1>
Parameter 3: <numeric 2>
Description
Value of the primary parameter
*1
.
Value of the secondary parameter
*1
.
Range
-99.9999E9 to 99.9999E9
-99.9999E9 to 99.9999E9
Initial value
0
0
Unit
Depends on the type of parameter. Depends on the type of parameter.
*1.Use the
[:SENSe]:CORRection:CKIT:STAN2:FORMat
command to specify the type
of parameter.
Содержание E4981A
Страница 2: ......
Страница 20: ...18 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 60: ...58 Chapter4 Preparation for Accurate Measurement Executing Correction Using the Cable Correction Function ...
Страница 88: ...86 Chapter7 Sorting Based on Measured Result Comparator Function Sample Program ...
Страница 162: ...160 Chapter9 Measurement Applications Sample Programs Detecting Error Occurrence using SRQ HT Basic ...
Страница 376: ...374 AppendixE Error Messages Warning Messages WARNING ...
Страница 382: ......
Страница 383: ...This information is subject to change without notice Keysight Technologies 2008 2014 Edition 2 August 2014 E4981 90001 ...