238
Chapter 10
Command Reference
E4981A SCPI Commands
[:SENSe]:CORRection:LOAD[:STATe]
Syntax
[:SENSe]:CORRection:LOAD[:STATe] {ON|OFF|1|0}
[:SENSe]:CORRection:LOAD[:STATe]?
Description
Turns ON/OFF the LOAD correction function.
With the LOAD correction set to ON, if you change the cable length (set with the
command) or frequency shift (set with the
command), the LOAD correction is automatically changed to OFF.
Parameters
Query response
{1|0}<newline><^END>
Related commands
:CALibration:CABLe[:LENGth] on page 174
:SYSTem:FSHift on page 276
[:SENSe]:CORRection:COLLect[:ACQuire] on page 233
Equivalent key
sequence
[
Meas Setup
] - CORRECTION - LOAD - ON/OFF
Description
ON or 1
Turns ON the LOAD correction.
OFF or 0 (initial value)
Turns OFF the LOAD correction.