74
Chapter 6
Reading Out Measured Result
Reading Out Measured Result
Reading out results of several measurements in batch (using data
buffer)
You can use the data buffer to temporarily store the results of several measurements and
then later read out these results in batch.
Data buffer types
There are three types of data buffers (buffers 1–3). Buffers 1 and 2 have the same function.
Buffer 3 has a different function as shown in the below table.
Feed position
Data fed into the buffer is stored sequentially in the order of measurement. When the data
is read out, the feed position returns to the beginning of the buffer and storage newly starts
from the position of the first data set (
).
The feed position also returns to the beginning of the buffer in the following case:
• When you set up the number of measurements to be fed into the data buffer.
Buffers 1 and 2
Buffer 3
Maximum number of measurements
that can be fed
200
1000
Data to be fed
for each
measurement
When the
comparator
function is OFF
3 data items: measurement
status, measured value of the
primary parameter or secondary
parameter
*1
, and comparator
sorting result
*2
(Independent of ON/OFF of the
comparator function)
3 data items: measurement
status, measured primary
parameter value, and measured
secondary parameter value
When the
comparator
function is ON
4 data items: measurement
status, measured primary
parameter value, measured
secondary parameter value, and
comparator sorting result
*1.Use the
:DATA:FEED:BUF1 command on page 194
to select the parameter to store
into the buffer from the primary or secondary parameters.
*2.When the comparator function is OFF, 11 is read out.
Содержание E4981A
Страница 2: ......
Страница 20: ...18 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 60: ...58 Chapter4 Preparation for Accurate Measurement Executing Correction Using the Cable Correction Function ...
Страница 88: ...86 Chapter7 Sorting Based on Measured Result Comparator Function Sample Program ...
Страница 162: ...160 Chapter9 Measurement Applications Sample Programs Detecting Error Occurrence using SRQ HT Basic ...
Страница 376: ...374 AppendixE Error Messages Warning Messages WARNING ...
Страница 382: ......
Страница 383: ...This information is subject to change without notice Keysight Technologies 2008 2014 Edition 2 August 2014 E4981 90001 ...