Chapter 6
73
Reading Out Measured Result
Reading Out Measured Result
6.
Re
ad
in
g
ou
t Me
as
ur
ed
Result
Figure 6-8
Behavior of /READY_FOR_TRIG signal at execution of
If the
command is executed when the trigger system is in the idle state, the
/READY_FOR_TRIG signal of the handler interface changes from HIGH to LOW after
completion of the internal processing following the reception of the command as shown in
.
If the trigger system continuous startup is ON, the /READY_FOR_TRIG signal changes
from HIGH to LOW (the trigger system transitions to the trigger wait state instead of the
idle state) after the completion of measurement and display update as shown in
As a result, it is not possible to determine whether the
been executed. Therefore, you need to turn OFF the trigger system continuous startup
(execute the
:INITiate:CONTinuous command on page 224
specifying OFF) in
advance.