Chapter 10
167
Command Reference
IEEE Common Commands
10
. Comm
and
Refe
ren
ce
*RST
Syntax
*RST
Description
Resets the instrument to the preset state.
The preset state is different from that when resetting is done using the
command. For details, refer to
Operation Manual, Appendix C “Initial Setting”
(No query)
Related commands
:SYSTem:PRESet on page 279
:INITiate:CONTinuous on page 224
Equivalent key
sequence
No equivalent key is available on the front panel.
*SAV
Syntax
*SAV <numeric>
Description
Saves the instrument setups to the register of the specified number on Flash (internal)
memory or USB pen drive (external memory). For details of saved instrument setups, refer
to
Operation Manual, Appendix C “Initial Setting”
(No query)
Parameters
If the specified parameter is out of the allowable setup range, an error occurs.
Related commands
Equivalent key
sequence
[
Save/Recall
] - No {1-9} - SAVE
<numeric>
Description
Specified number
Range
0 to 9 (Internal Flash Memory)
10 to 19 (External USB Memory)
Resolution
1
Содержание E4981A
Страница 2: ......
Страница 20: ...18 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 60: ...58 Chapter4 Preparation for Accurate Measurement Executing Correction Using the Cable Correction Function ...
Страница 88: ...86 Chapter7 Sorting Based on Measured Result Comparator Function Sample Program ...
Страница 162: ...160 Chapter9 Measurement Applications Sample Programs Detecting Error Occurrence using SRQ HT Basic ...
Страница 376: ...374 AppendixE Error Messages Warning Messages WARNING ...
Страница 382: ......
Страница 383: ...This information is subject to change without notice Keysight Technologies 2008 2014 Edition 2 August 2014 E4981 90001 ...