Chapter 10
Command Reference
E4981A SCPI Commands
E4981A SCPI Commands
This section describes the SCPI commands of the E4981A.
:ABORt
Syntax
:ABORt
Description
Resets the trigger system and places the trigger sequence in the idle state.
If the trigger system is set to start up successively (ON is specified with the
command), the trigger system starts up immediately after the
transition to the idle state.
For details on the trigger system, refer to
. (No query)
Related commands
:INITiate[:IMMediate] on page 224
:INITiate:CONTinuous on page 224
Equivalent key
sequence
No equivalent key is available on the front panel.
:CALibration:CABLe:CORRection:CLEar
Syntax
:CALibration:CABLe:CORRection:CLEar <numeric>
Description
Clears the cable correction data. (No query)
Parameters
Related commands
:CALibration:CABLe:CORRection:COLLect[:ACQuire]:LOAD on page 171
:CALibration:CABLe:CORRection:COLLect[:ACQuire]:OPEN on page 171
:CALibration:CABLe:CORRection:COLLect[:ACQuire]:REFerence on page 172
:CALibration:CABLe:CORRection:SAVE on page 172
:CALibration:CABLe:CORRection:STATe? on page 173
Equivalent key
sequence
[
System
] - CABLE CORR - 1m|2m - CLEAR
<numeric>
Description
Cable Length (in meters)
Range
1 or 2
Содержание E4981A
Страница 2: ......
Страница 20: ...18 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 60: ...58 Chapter4 Preparation for Accurate Measurement Executing Correction Using the Cable Correction Function ...
Страница 88: ...86 Chapter7 Sorting Based on Measured Result Comparator Function Sample Program ...
Страница 162: ...160 Chapter9 Measurement Applications Sample Programs Detecting Error Occurrence using SRQ HT Basic ...
Страница 376: ...374 AppendixE Error Messages Warning Messages WARNING ...
Страница 382: ......
Страница 383: ...This information is subject to change without notice Keysight Technologies 2008 2014 Edition 2 August 2014 E4981 90001 ...