DriverLINX Calibration and Test Utilities
Equipment Required
You need the following equipment to calibrate a DDA-08/16 board:
●
A digital multimeter accurate to 6
½
digits on its ±10VDC range, such
as the Keithley DMM Model 196.
●
An STA-U screw terminal accessory and C-1800 cable, and STP-37
screw terminal panel and C-1800 cable, or an STC-37 screw terminal
connector.
●
A small screwdriver.
DriverLINX Calibration Utility will guide you through the calibration
procedure. Before calibration, specify the following parameters in the
setup panel to get the correct instructions:
●
Logical Device
—
Board’s device number, model, and address.
●
Accessory
—
Connection method used to connect the board to the
calibration stimulus.
●
Shorted channel
—
Input channel to be “shorted” high to low.
●
Voltage Channel
—
Input channel to use to apply the various
calibration voltage levels.
●
Calibration range
—
Input range to be calibrated.
Note:
Do not attempt to calibrate a DDA-08/16 board unless your
equipment is of the required accuracy.