If your DDA-08/16 board is not operating properly, use the information in
this section to help you isolate the problem. If the problem appears
serious enough to require technical support, refer to page 7-6 for
information on how to contact an applications engineer.
Identifying Symptoms and Possible Causes
Table 7-1 lists general symptoms and possible solutions for problems with
DDA-08/16 boards. If your board is not operating properly after using
this information, refer to page 7-6 for instructions on getting technical
support.
Table 7-1. Troubleshooting Information
Symptom
Possible Cause
Possible Solution
Board does not respond
Base address is unacceptable.
Make sure that the base address
specified in the configuration file
matches the setting of the base
address switch block on the
board. Make sure that no other
system resource is using any of
the eight memory locations
starting at the specified base
address. Reconfigure the base
address, if necessary. Refer to
page 3-10 for instructions.