![background image](http://html1.mh-extra.com/html/keithley/dda-08/dda-08_user-manual_661637042.webp)
3-11
Figure 3-2 illustrates the setting for a base address of 300h (768 decimal).
Switches 1 and 2 are in the OFF position (200h + 100h = 300h;
512 + 256 = 768); all the other switches (3, 4, 5, 6, and 7) are in the ON
position.
Figure 3-2. Setting the Base Address
Table 3-2. Base Address Switches
Switch
Value When Switch is OFF
Hexadecimal
Decimal
1
200
512
2
100
256
3
80
128
4
40
64
5
20
32
6
10
16
7
8
8
1
2
3
4
5
6
7
O
N