74
3.6 Replacing the boards
Following is a list what should be checked/calibrated after
replacing certain parts on the system. In addition, Quality
Check (QC) images of all modalities should be taken and
inspected every time after replacing any part on the
system.
EG200 X-ray Interface Board (rotating unit)
1. Preheat calibration
2. mA calibration
EG200 X-ray Interface Board (ceph head)
1. Check operation of all cephalostat patient positioning
panel buttons
2. Soft tissue filter calibration
3. Take lateral ceph images to test soft tissue filter
operation (by moving nasion support)
EG400 Panel PSU (rotator unit)
1. Pan pixel calibration
2. 3D pixel calibration
EG400 Panel PSU (ceph)
1. Ceph pixel calibration
R3220 CPU Board (in the rotator unit)
1. Upgrade the unit’s embedded software. Use automatic
update to ensure all software components are compat-
ible. The R3220 CPU board must be separately updat-
ed before a full update is done if the replacement board
is not preprogrammed with R3220 Main CPU software.
Instructions on how to do this are included in the soft-
ware release package.
2. Perform full calibration of the unit (starting with preheat
and mA calibrations)
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X ray System Service Manual ENGLISH 216508 rev 1 0 805 5127...
Страница 2: ......
Страница 4: ......
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 28: ...2 Unit description 20 2 6 Unit movements Panoramic unit movements R C J D...
Страница 31: ...2 Unit description 23...
Страница 32: ...2 Unit description 24...
Страница 37: ...3 Electrical description 29 Circuit boards Cephalometric unit...
Страница 54: ...46...
Страница 87: ...79...
Страница 88: ...80...
Страница 96: ...4 Firmware description 88...
Страница 101: ...5 Hardware operation 216508 rev 1 NEW KAVO 93 5 6 Z movement 2...
Страница 105: ...5 Hardware operation 216508 rev 1 NEW KAVO 97 5 10 Secondary collimator movement 2...
Страница 106: ...5 Hardware operation 98 NEW KAVO 216508 rev 1 5 11 Ceph sensor movement 2...
Страница 107: ...5 Hardware operation 216508 rev 1 NEW KAVO 99 5 12 Imaging chain 5 12 1 3D imaging chain...
Страница 108: ...5 Hardware operation 100 NEW KAVO 216508 rev 1 5 12 2 Panoramic imaging chain...
Страница 109: ...5 Hardware operation 216508 rev 1 NEW KAVO 101 5 12 3 Cephalometric imaging chain...
Страница 110: ...5 Hardware operation 102 NEW KAVO 216508 rev 1 5 12 4 Exposure control...
Страница 112: ...5 Hardware operation 104 NEW KAVO 216508 rev 1...
Страница 122: ...6 The s2terminal and service functions 114...
Страница 192: ...8 Maintenance procedure 184...