6 The s2terminal and service functions
112
------------------------------------------------------------------
Note!
This list may vary from unit to unit.
3. To open help instructions for a service command key in
help
, space, the name of the service command, for ex-
ample
motortest
, and then press
ENTER
.
------------------------------------------------------------------
------------------------------------------------------------------
Help instruction for that service command will appear.
------------------------------------------------------------------
------------------------------------------------------------------
4. To switch to demo mode use the service command
demomode
.
------------------------------------------------------------------
------------------------------------------------------------------
sets the unit into demonstration mode.
- x-rays are disabled
- no need to have ethernet connection to pc to activate
gui
log
Displays the log entries.
login
Displays unit’s firmware version informa-
tion
logsign
Adds a comment to the log
memtest
Performs a memory test on image memo-
ry area.
motortest
Runs a motor test.
patient
Starts fake study without using driver
posprint
Prints position information of selected
motor.
panelpower
Sets the panel on/off.
readadc
Read adc channels
status
Displays unit’s status information.
systemstatus
Displays status of all optos, micro switch-
es and encoders.
unithandedness
Sets the unit’s handedness.
vaccalib
Calibrates VAC value using real voltage
value.
help motortest
help motortest
Usage:
motortest [rounds] [motor]
motortest [rounds]
Drives all or specific motor from limit to limit.
demomode 1
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X ray System Service Manual ENGLISH 216508 rev 1 0 805 5127...
Страница 2: ......
Страница 4: ......
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 28: ...2 Unit description 20 2 6 Unit movements Panoramic unit movements R C J D...
Страница 31: ...2 Unit description 23...
Страница 32: ...2 Unit description 24...
Страница 37: ...3 Electrical description 29 Circuit boards Cephalometric unit...
Страница 54: ...46...
Страница 87: ...79...
Страница 88: ...80...
Страница 96: ...4 Firmware description 88...
Страница 101: ...5 Hardware operation 216508 rev 1 NEW KAVO 93 5 6 Z movement 2...
Страница 105: ...5 Hardware operation 216508 rev 1 NEW KAVO 97 5 10 Secondary collimator movement 2...
Страница 106: ...5 Hardware operation 98 NEW KAVO 216508 rev 1 5 11 Ceph sensor movement 2...
Страница 107: ...5 Hardware operation 216508 rev 1 NEW KAVO 99 5 12 Imaging chain 5 12 1 3D imaging chain...
Страница 108: ...5 Hardware operation 100 NEW KAVO 216508 rev 1 5 12 2 Panoramic imaging chain...
Страница 109: ...5 Hardware operation 216508 rev 1 NEW KAVO 101 5 12 3 Cephalometric imaging chain...
Страница 110: ...5 Hardware operation 102 NEW KAVO 216508 rev 1 5 12 4 Exposure control...
Страница 112: ...5 Hardware operation 104 NEW KAVO 216508 rev 1...
Страница 122: ...6 The s2terminal and service functions 114...
Страница 192: ...8 Maintenance procedure 184...