7 Error codes and troubleshooting
154
Head support is in wrong position
<ID, KAVO, GDX only>
CAUSE A
The head support position is not valid for the select-
ed imaging program.
SOLUTION A
Lift the head support to the upmost position.
CAUSE B
The head support position detection is not detected
correctly.
SOLUTION B
Check that the position detection micro switch is
activated / deactivated properly when the head
support is lifted / lowered. Check the cable between
the micro switch and EG200/JG202. Check the se-
rial data line between EG200/JG205 and R3210/
J3213. Replace micro switch, cables or EG200 if
needed.
Image processing error notifications
CAUSE A
The image was not transferred to the PC correctly,
or processing failed.
SOLUTION A
If the current exam was closed, open a new exam
for the same patient in the image acquisition work-
station. Press the “Retrieve last image” button in
the GUI to have the image re-transferred.
All needed calibrations are not yet done
CAUSE A
One or more calibrations must be completed before
imaging is allowed for all modalities.
SOLUTION A
Switch to the service / quality assurance view in the
GUI to find out which calibration is not done. Com-
plete all calibrations marked as not done.
The sensor was moved from its correct position
CAUSE A
The sensor was moved (rotated) out of position
during exposure.
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X ray System Service Manual ENGLISH 216508 rev 1 0 805 5127...
Страница 2: ......
Страница 4: ......
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 28: ...2 Unit description 20 2 6 Unit movements Panoramic unit movements R C J D...
Страница 31: ...2 Unit description 23...
Страница 32: ...2 Unit description 24...
Страница 37: ...3 Electrical description 29 Circuit boards Cephalometric unit...
Страница 54: ...46...
Страница 87: ...79...
Страница 88: ...80...
Страница 96: ...4 Firmware description 88...
Страница 101: ...5 Hardware operation 216508 rev 1 NEW KAVO 93 5 6 Z movement 2...
Страница 105: ...5 Hardware operation 216508 rev 1 NEW KAVO 97 5 10 Secondary collimator movement 2...
Страница 106: ...5 Hardware operation 98 NEW KAVO 216508 rev 1 5 11 Ceph sensor movement 2...
Страница 107: ...5 Hardware operation 216508 rev 1 NEW KAVO 99 5 12 Imaging chain 5 12 1 3D imaging chain...
Страница 108: ...5 Hardware operation 100 NEW KAVO 216508 rev 1 5 12 2 Panoramic imaging chain...
Страница 109: ...5 Hardware operation 216508 rev 1 NEW KAVO 101 5 12 3 Cephalometric imaging chain...
Страница 110: ...5 Hardware operation 102 NEW KAVO 216508 rev 1 5 12 4 Exposure control...
Страница 112: ...5 Hardware operation 104 NEW KAVO 216508 rev 1...
Страница 122: ...6 The s2terminal and service functions 114...
Страница 192: ...8 Maintenance procedure 184...