7 Error codes and troubleshooting
134
7.4.22
Error 37 Movement of R (Rotation)
motor failed
CAUSE A
The rotator is not moving correctly.
SOLUTION A
The movement not working correctly may be due to
a fault in the
• motor,
• motor cable,
• R5100 Stepper driver module,
• cable between the stepper driver module and the
CPU board or the
• R3300 IO board.
Use service command “drivemotor r manual” to test
whether the motor is working properly. If the chin
support is moving, the problem is probably in posi-
tion detection (see CAUSE B). Otherwise one of
the components listed above is faulty. Continue
with inspecting cables and connectors. Replace
parts one by one if needed.
CAUSE B
Position detection showing incorrect values.
SOLUTION B
Check that cable 204808 is properly connected to
Rotator opto board EB100. Reconnect or replace
the cable and/or opto board if needed. The opera-
tion of two of the opto switches can be checked with
service program “systemstatus” (signals “R outer
opto” and “R inner opto”)
CAUSE C
Faulty R3300 IO board.
SOLUTION C
Replace the R3300 IO board.
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X ray System Service Manual ENGLISH 216508 rev 1 0 805 5127...
Страница 2: ......
Страница 4: ......
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 28: ...2 Unit description 20 2 6 Unit movements Panoramic unit movements R C J D...
Страница 31: ...2 Unit description 23...
Страница 32: ...2 Unit description 24...
Страница 37: ...3 Electrical description 29 Circuit boards Cephalometric unit...
Страница 54: ...46...
Страница 87: ...79...
Страница 88: ...80...
Страница 96: ...4 Firmware description 88...
Страница 101: ...5 Hardware operation 216508 rev 1 NEW KAVO 93 5 6 Z movement 2...
Страница 105: ...5 Hardware operation 216508 rev 1 NEW KAVO 97 5 10 Secondary collimator movement 2...
Страница 106: ...5 Hardware operation 98 NEW KAVO 216508 rev 1 5 11 Ceph sensor movement 2...
Страница 107: ...5 Hardware operation 216508 rev 1 NEW KAVO 99 5 12 Imaging chain 5 12 1 3D imaging chain...
Страница 108: ...5 Hardware operation 100 NEW KAVO 216508 rev 1 5 12 2 Panoramic imaging chain...
Страница 109: ...5 Hardware operation 216508 rev 1 NEW KAVO 101 5 12 3 Cephalometric imaging chain...
Страница 110: ...5 Hardware operation 102 NEW KAVO 216508 rev 1 5 12 4 Exposure control...
Страница 112: ...5 Hardware operation 104 NEW KAVO 216508 rev 1...
Страница 122: ...6 The s2terminal and service functions 114...
Страница 192: ...8 Maintenance procedure 184...