7 Error codes and troubleshooting
158
7.6.2
Power distribution (mains)
7.6.2.1
Nothing happens when the X-ray unit is
switched on
•
Check that the mains cord is connected to the wall
socket.
•
Check that there is mains power reaching the wall
socket.
•
Check Fuses F1 and F2.
•
Open the front cover at the lower end of the car-
riage and check that the voltage indicator lamp on
EA200 is on (or measure voltage from connector
JA203).
7.6.2.2
No X-rays
•
Check that lamp (LA1) on R3710 (generator pow-
er) is on.
If not, plug it in.
If not, call your local electrician to fix the problem.
Check that correct voltage range is selected by
checking the position of the voltage selector
jumper J1 on the EA200 board.
Replace blown fuses.
If EA200 is receiving voltage then measure the
secondary voltages from the low power supply.
If any of voltages are missing when each
of the low voltage connectors are un-
plugged, replace transformer.
If voltages are OK, when each low volt-
age connector is un-plugged, but not OK
when any of connectors is connected,
then follow trouble shooting for the LOW
voltage side.
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X ray System Service Manual ENGLISH 216508 rev 1 0 805 5127...
Страница 2: ......
Страница 4: ......
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 28: ...2 Unit description 20 2 6 Unit movements Panoramic unit movements R C J D...
Страница 31: ...2 Unit description 23...
Страница 32: ...2 Unit description 24...
Страница 37: ...3 Electrical description 29 Circuit boards Cephalometric unit...
Страница 54: ...46...
Страница 87: ...79...
Страница 88: ...80...
Страница 96: ...4 Firmware description 88...
Страница 101: ...5 Hardware operation 216508 rev 1 NEW KAVO 93 5 6 Z movement 2...
Страница 105: ...5 Hardware operation 216508 rev 1 NEW KAVO 97 5 10 Secondary collimator movement 2...
Страница 106: ...5 Hardware operation 98 NEW KAVO 216508 rev 1 5 11 Ceph sensor movement 2...
Страница 107: ...5 Hardware operation 216508 rev 1 NEW KAVO 99 5 12 Imaging chain 5 12 1 3D imaging chain...
Страница 108: ...5 Hardware operation 100 NEW KAVO 216508 rev 1 5 12 2 Panoramic imaging chain...
Страница 109: ...5 Hardware operation 216508 rev 1 NEW KAVO 101 5 12 3 Cephalometric imaging chain...
Страница 110: ...5 Hardware operation 102 NEW KAVO 216508 rev 1 5 12 4 Exposure control...
Страница 112: ...5 Hardware operation 104 NEW KAVO 216508 rev 1...
Страница 122: ...6 The s2terminal and service functions 114...
Страница 192: ...8 Maintenance procedure 184...