7 Error codes and troubleshooting
116
s2terminal command:
- Downloading raw images - xi, these images can be
use for troubleshooting.
Refer to section,
The s2terminal and service functions
.
Checking circuit boards
Circuit boards cannot be repaired in the field. Some boards
have fuses that can be replaced. However, if a board is
faulty, replace it.
On most of the circuit boards there are indicator LEDs (D)
and test pins (TP), that allow the operation of the board to
be checked and/or tested. The LEDs and test pins for each
circuit board are described in the section
Electrical
Description
. Use a digital multimeter (DMM) when
checking boards.
DANGER HIGH VOLTAGE:
R3400 IO2, R3710 X-ray power supply, R3810 X-ray
generator and R3900 Mains connector include high voltage
indicator lamps (Glim). When a high voltage lamp is on it
indicates that high voltage is present. DO NOT touch the
board until the high voltage has dissapated and the lamp
has gone out.
When handling circuit boards take all necessary
precautions to prevent the build up of static electricity and
possible static damage to the boards.
Checking cables and connectors
Visually check cables for mechanical damage, cuts,
damaged insulation and bad twists. If a cable is damaged
in any way replace it.
If there is no obvious mechanical damage to a cable but
you think that it may be faulty, use a digital multimeter
(DMM) to check the resistance of the different wires within
the cable. An undamaged wire will have close to no
resistance (>0 ohm), a damaged wire will have a high
resistance value.
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X ray System Service Manual ENGLISH 216508 rev 1 0 805 5127...
Страница 2: ......
Страница 4: ......
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 28: ...2 Unit description 20 2 6 Unit movements Panoramic unit movements R C J D...
Страница 31: ...2 Unit description 23...
Страница 32: ...2 Unit description 24...
Страница 37: ...3 Electrical description 29 Circuit boards Cephalometric unit...
Страница 54: ...46...
Страница 87: ...79...
Страница 88: ...80...
Страница 96: ...4 Firmware description 88...
Страница 101: ...5 Hardware operation 216508 rev 1 NEW KAVO 93 5 6 Z movement 2...
Страница 105: ...5 Hardware operation 216508 rev 1 NEW KAVO 97 5 10 Secondary collimator movement 2...
Страница 106: ...5 Hardware operation 98 NEW KAVO 216508 rev 1 5 11 Ceph sensor movement 2...
Страница 107: ...5 Hardware operation 216508 rev 1 NEW KAVO 99 5 12 Imaging chain 5 12 1 3D imaging chain...
Страница 108: ...5 Hardware operation 100 NEW KAVO 216508 rev 1 5 12 2 Panoramic imaging chain...
Страница 109: ...5 Hardware operation 216508 rev 1 NEW KAVO 101 5 12 3 Cephalometric imaging chain...
Страница 110: ...5 Hardware operation 102 NEW KAVO 216508 rev 1 5 12 4 Exposure control...
Страница 112: ...5 Hardware operation 104 NEW KAVO 216508 rev 1...
Страница 122: ...6 The s2terminal and service functions 114...
Страница 192: ...8 Maintenance procedure 184...