7 Error codes and troubleshooting
141
7.4.31
Error 46 Emergency stop button is
pressed
CAUSE A
The emergency stop button is activated.
SOLUTION A
Release emergency stop button and dismiss the
error message. If the unit is equipped with a ceph-
alostat, emergency may also be activated from the
cephalostat.
CAUSE B
The emergency stop switch is not working properly.
The switch shall open the circuit when pressed and
make a short circuit otherwise.
SOLUTION B
Check the switch operation with a multimeter and
replace if faulty. If the unit is equipped with a ceph-
alostat, check also the emergency stop button lo-
cated in the bottom of the cephalostat.
CAUSE C
The loopback connector shorting the cephalostat
emergency stop for non-ceph units is missing (shall
be connected to JE006 in the top of the pillar).
SOLUTION C
Attach the loopback connector, or if it’s present,
check that the emergency stop signal is passing
through the connector properly.
CAUSE D
Emergency stop switch cabling is faulty.
SOLUTION D
Short the pins in JA303 and JA304 in the EA300
Extension board. If the error is no longer present,
the cables are faulty and need to be repaired/re-
placed. If shorting the pins in the J304 removes the
error, the problem is related to the cephalostat (ca-
bles, emergency switch, connectors). Check and
repair/replace cables if needed.
CAUSE E
EA300 extension board or R3400 IO board is
faulty.
SOLUTION E
If the error is still present with the pins in JA303 and
JA304 shorted, the error is in the EA300 or R3400
board. Replace both boards.
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X ray System Service Manual ENGLISH 216508 rev 1 0 805 5127...
Страница 2: ......
Страница 4: ......
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 28: ...2 Unit description 20 2 6 Unit movements Panoramic unit movements R C J D...
Страница 31: ...2 Unit description 23...
Страница 32: ...2 Unit description 24...
Страница 37: ...3 Electrical description 29 Circuit boards Cephalometric unit...
Страница 54: ...46...
Страница 87: ...79...
Страница 88: ...80...
Страница 96: ...4 Firmware description 88...
Страница 101: ...5 Hardware operation 216508 rev 1 NEW KAVO 93 5 6 Z movement 2...
Страница 105: ...5 Hardware operation 216508 rev 1 NEW KAVO 97 5 10 Secondary collimator movement 2...
Страница 106: ...5 Hardware operation 98 NEW KAVO 216508 rev 1 5 11 Ceph sensor movement 2...
Страница 107: ...5 Hardware operation 216508 rev 1 NEW KAVO 99 5 12 Imaging chain 5 12 1 3D imaging chain...
Страница 108: ...5 Hardware operation 100 NEW KAVO 216508 rev 1 5 12 2 Panoramic imaging chain...
Страница 109: ...5 Hardware operation 216508 rev 1 NEW KAVO 101 5 12 3 Cephalometric imaging chain...
Страница 110: ...5 Hardware operation 102 NEW KAVO 216508 rev 1 5 12 4 Exposure control...
Страница 112: ...5 Hardware operation 104 NEW KAVO 216508 rev 1...
Страница 122: ...6 The s2terminal and service functions 114...
Страница 192: ...8 Maintenance procedure 184...