7 Error codes and troubleshooting
149
7.5 Warnings and notifications
Please press any button to complete unit start up
CAUSE A
The unit needs the operators acknowledgement to
start initialization movements.
SOLUTION A
Press any button on the patient positioning panel to
start movement initialization. Wait for movements
to complete.
Please wait for X-Ray tube to cool down
CAUSE A
The X-ray tube needs to cool down before next ex-
posure is started.
SOLUTION A
Wait for the X-ray tube to cool down. If the cool
down time is long (900 seconds), the cool down
time is caused by a hardware protection mecha-
nism. Power off the unit and let it cool down for 15
minutes. If the error still appears, check possible
causes below.
CAUSE B
Loose or faulty generator temperature feedback
cables.
SOLUTION B
Check that the following cables are intact and prop-
erly connected:
• 205032 between EG200/JG215 and R3810/
J3807.
• Tubehead filament and feedback signals cable
connected to J3809 in R3810.
Reconnect or replace if needed.
CAUSE C
Faulty EG200 XRAY IF board.
SOLUTION C
Replace the EG200 XRAY IF board.
CAUSE D
Faulty tubehead thermal switch or thermal switch
connection.
SOLUTION D
Replace the tubehead.
CAUSE E
Generator temperature monitoring faulty or miscal-
ibrated.
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X ray System Service Manual ENGLISH 216508 rev 1 0 805 5127...
Страница 2: ......
Страница 4: ......
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 28: ...2 Unit description 20 2 6 Unit movements Panoramic unit movements R C J D...
Страница 31: ...2 Unit description 23...
Страница 32: ...2 Unit description 24...
Страница 37: ...3 Electrical description 29 Circuit boards Cephalometric unit...
Страница 54: ...46...
Страница 87: ...79...
Страница 88: ...80...
Страница 96: ...4 Firmware description 88...
Страница 101: ...5 Hardware operation 216508 rev 1 NEW KAVO 93 5 6 Z movement 2...
Страница 105: ...5 Hardware operation 216508 rev 1 NEW KAVO 97 5 10 Secondary collimator movement 2...
Страница 106: ...5 Hardware operation 98 NEW KAVO 216508 rev 1 5 11 Ceph sensor movement 2...
Страница 107: ...5 Hardware operation 216508 rev 1 NEW KAVO 99 5 12 Imaging chain 5 12 1 3D imaging chain...
Страница 108: ...5 Hardware operation 100 NEW KAVO 216508 rev 1 5 12 2 Panoramic imaging chain...
Страница 109: ...5 Hardware operation 216508 rev 1 NEW KAVO 101 5 12 3 Cephalometric imaging chain...
Страница 110: ...5 Hardware operation 102 NEW KAVO 216508 rev 1 5 12 4 Exposure control...
Страница 112: ...5 Hardware operation 104 NEW KAVO 216508 rev 1...
Страница 122: ...6 The s2terminal and service functions 114...
Страница 192: ...8 Maintenance procedure 184...