6 The s2terminal and service functions
109
6.3 Unit state commands
The unit state commands, allow the states of the unit to be
checked and changed.
Note!
Some of unit state commands can also be accessed
when the unit is in service mode.
status r
This command shows the current state of the unit.
The states are:
-
idle
state. The unit is ready to start an examination
and is waiting for image capture, from the dental
imaging software, to be activated (a message re-
questing this will appear on the touch screen dis-
play).
-
select
state. The imaging program and technique
factors can be selected. When the
RETURN
key is
pressed, to drive the rotating unit to the ready posi-
tion, the unit will enter the next state, ready.
-
ready
state. The unit is ready to take an exposure.
The text
READY
will appear on the touch screen
display.
-
exposure
state. The unit will enter this state when
the exposure switch is pressed and held down. The
unit will enter the next state when the exposure
button is released after the exposure is complete.
-
transfer
_check state. The image to being trans-
ferred from the unit to the PC. Note that image
transfer already starts during exposure.
-
service
state. The unit is accepts the service com-
mands.
-
error
state. The unit will enter this state if it mal-
functions. An error message will appear on the unit
touch screen display and the s2terminal if it is con-
nected.
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X ray System Service Manual ENGLISH 216508 rev 1 0 805 5127...
Страница 2: ......
Страница 4: ......
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 28: ...2 Unit description 20 2 6 Unit movements Panoramic unit movements R C J D...
Страница 31: ...2 Unit description 23...
Страница 32: ...2 Unit description 24...
Страница 37: ...3 Electrical description 29 Circuit boards Cephalometric unit...
Страница 54: ...46...
Страница 87: ...79...
Страница 88: ...80...
Страница 96: ...4 Firmware description 88...
Страница 101: ...5 Hardware operation 216508 rev 1 NEW KAVO 93 5 6 Z movement 2...
Страница 105: ...5 Hardware operation 216508 rev 1 NEW KAVO 97 5 10 Secondary collimator movement 2...
Страница 106: ...5 Hardware operation 98 NEW KAVO 216508 rev 1 5 11 Ceph sensor movement 2...
Страница 107: ...5 Hardware operation 216508 rev 1 NEW KAVO 99 5 12 Imaging chain 5 12 1 3D imaging chain...
Страница 108: ...5 Hardware operation 100 NEW KAVO 216508 rev 1 5 12 2 Panoramic imaging chain...
Страница 109: ...5 Hardware operation 216508 rev 1 NEW KAVO 101 5 12 3 Cephalometric imaging chain...
Страница 110: ...5 Hardware operation 102 NEW KAVO 216508 rev 1 5 12 4 Exposure control...
Страница 112: ...5 Hardware operation 104 NEW KAVO 216508 rev 1...
Страница 122: ...6 The s2terminal and service functions 114...
Страница 192: ...8 Maintenance procedure 184...