8 Maintenance procedure
182
unit. The resistance must be below 0.2 Ohm.
8. Reattach the covers and connect the unit the mains pow-
er supply.
9. Check that all covers are correctly installed and in good
condition. Also check that all labels are attached to the
init and that they are legible.
10.Perform a motor movements using the s2terminal ser-
vice command “autotest”. Make sure the motors operat-
ed smoothly and without noise.
11.Perform all calibration and quality check programs.
12.Clear service indicator using s2terminal command “indi-
cator”.
Motors and movement mechanisms
Perform a check of the motors and movement mechanisms
before any covers are removed to survey the need for
service.
Check the X, Y, N and rotational movements for slack by
powering on the unit and making it ready for exposure in
the panoramic light service program. The rotator shall stay
fi rmly in position. No noticeable slack shall be present in X,
Y and N directions. For the rotational movement some
minor slack can be allowed, as long as the movement
centers itself to the same position after being lightly pushed
out of position. Tighten the rotational movement belts
(2 pcs) if needed.
Perform a motorized movement test using s2terminal
service command “motortest”. Check that all movements
are done smoothly and without noise. If any problems are
encountered, service the movement mechanism(s) as
needed. Clean away dirt and apply new lubrication. Fasten
loose parts and exchange worn out ones.
Mains power supply cable
Power off the unit and disconnect it from the mains power
supply. Check the condition of the main power supply
cable. Replace the cable if damaged.
Tubehead
Remove the covers around the tubehead. Make sure that
no oil has leaked from the tubehead. Replace the tubehead
if leakages are encountered.
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X ray System Service Manual ENGLISH 216508 rev 1 0 805 5127...
Страница 2: ......
Страница 4: ......
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 28: ...2 Unit description 20 2 6 Unit movements Panoramic unit movements R C J D...
Страница 31: ...2 Unit description 23...
Страница 32: ...2 Unit description 24...
Страница 37: ...3 Electrical description 29 Circuit boards Cephalometric unit...
Страница 54: ...46...
Страница 87: ...79...
Страница 88: ...80...
Страница 96: ...4 Firmware description 88...
Страница 101: ...5 Hardware operation 216508 rev 1 NEW KAVO 93 5 6 Z movement 2...
Страница 105: ...5 Hardware operation 216508 rev 1 NEW KAVO 97 5 10 Secondary collimator movement 2...
Страница 106: ...5 Hardware operation 98 NEW KAVO 216508 rev 1 5 11 Ceph sensor movement 2...
Страница 107: ...5 Hardware operation 216508 rev 1 NEW KAVO 99 5 12 Imaging chain 5 12 1 3D imaging chain...
Страница 108: ...5 Hardware operation 100 NEW KAVO 216508 rev 1 5 12 2 Panoramic imaging chain...
Страница 109: ...5 Hardware operation 216508 rev 1 NEW KAVO 101 5 12 3 Cephalometric imaging chain...
Страница 110: ...5 Hardware operation 102 NEW KAVO 216508 rev 1 5 12 4 Exposure control...
Страница 112: ...5 Hardware operation 104 NEW KAVO 216508 rev 1...
Страница 122: ...6 The s2terminal and service functions 114...
Страница 192: ...8 Maintenance procedure 184...