
7 Error codes and troubleshooting
125
7.4.5
Error 12 R3400 IO2 board is not
functioning
7.4.6
Error 13 Configuration data was lost
CAUSE A
Communication between the main CPU and R3400
is not working properly.
SOLUTION A
• The Ethernet cable between the R3400 and the
internal Ethernet switch may be faulty. Try with a
spare cable and replace the cable if needed.
• If also errors 9 is shown, check the cable between
the main CPU and the internal Ethernet switch. Try
with a spare cable and replace the cable if needed.
• Check the internal Ethernet switch and replace if
needed:
- Appropriate leds are lit: each port having a
cable connected should have the link led
on.
- The specified voltage level is present in
power connector of the switch.
CAUSE B
The R3400 IO board software is incompatible with
the main CPU software. State commanded for ex-
ternal warning light is not matching feedback value.
SOLUTION B
Update the unit’s embedded software. Verify that
the versions listed by the s2terminal command
“login” match versions.txt in the software update
package.
CAUSE C
The R3400 IO board is faulty and therefore not re-
sponding.
SOLUTION C
Replace the board.
CAUSE A
Main CPU flash memory corrupted.
SOLUTION A
Recalibrate the unit. Replace the main CPU board
if the problem reoccurs.
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X ray System Service Manual ENGLISH 216508 rev 1 0 805 5127...
Страница 2: ......
Страница 4: ......
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 28: ...2 Unit description 20 2 6 Unit movements Panoramic unit movements R C J D...
Страница 31: ...2 Unit description 23...
Страница 32: ...2 Unit description 24...
Страница 37: ...3 Electrical description 29 Circuit boards Cephalometric unit...
Страница 54: ...46...
Страница 87: ...79...
Страница 88: ...80...
Страница 96: ...4 Firmware description 88...
Страница 101: ...5 Hardware operation 216508 rev 1 NEW KAVO 93 5 6 Z movement 2...
Страница 105: ...5 Hardware operation 216508 rev 1 NEW KAVO 97 5 10 Secondary collimator movement 2...
Страница 106: ...5 Hardware operation 98 NEW KAVO 216508 rev 1 5 11 Ceph sensor movement 2...
Страница 107: ...5 Hardware operation 216508 rev 1 NEW KAVO 99 5 12 Imaging chain 5 12 1 3D imaging chain...
Страница 108: ...5 Hardware operation 100 NEW KAVO 216508 rev 1 5 12 2 Panoramic imaging chain...
Страница 109: ...5 Hardware operation 216508 rev 1 NEW KAVO 101 5 12 3 Cephalometric imaging chain...
Страница 110: ...5 Hardware operation 102 NEW KAVO 216508 rev 1 5 12 4 Exposure control...
Страница 112: ...5 Hardware operation 104 NEW KAVO 216508 rev 1...
Страница 122: ...6 The s2terminal and service functions 114...
Страница 192: ...8 Maintenance procedure 184...