7 Error codes and troubleshooting
127
7.4.9
Error 23 Configuration data is invalid
7.4.10
Error 24 Voltage (kV) failure in R3810
X-ray generator
7.4.11
Error 25 Current (mA) failure in R3810
X-ray generator
CAUSE A
Unit configuration data stored in the main CPU
board is missing or corrupted.
SOLUTION A
Recalibrate the unit. Replace the main CPU board
if the problem reoccurs.
The kV feedback signal from the generator is out of
range. See also Error 6.
CAUSE A
The exposure enable signal is not reaching the X-
ray generator.
SOLUTION A
Ensure that the exposure enable signal cable
(205467) is properly connected to EA200/JA206
and EG200/JG206.
The mA feedback signal from the generator is out
of range. See also Error 6.
CAUSE A
The exposure enable signal is not reaching the X-
ray generator.
SOLUTION A
Ensure that the exposure enable signal cable
(205467) is properly connected to EA200/JA206
and EG200/JG206.
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X ray System Service Manual ENGLISH 216508 rev 1 0 805 5127...
Страница 2: ......
Страница 4: ......
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 28: ...2 Unit description 20 2 6 Unit movements Panoramic unit movements R C J D...
Страница 31: ...2 Unit description 23...
Страница 32: ...2 Unit description 24...
Страница 37: ...3 Electrical description 29 Circuit boards Cephalometric unit...
Страница 54: ...46...
Страница 87: ...79...
Страница 88: ...80...
Страница 96: ...4 Firmware description 88...
Страница 101: ...5 Hardware operation 216508 rev 1 NEW KAVO 93 5 6 Z movement 2...
Страница 105: ...5 Hardware operation 216508 rev 1 NEW KAVO 97 5 10 Secondary collimator movement 2...
Страница 106: ...5 Hardware operation 98 NEW KAVO 216508 rev 1 5 11 Ceph sensor movement 2...
Страница 107: ...5 Hardware operation 216508 rev 1 NEW KAVO 99 5 12 Imaging chain 5 12 1 3D imaging chain...
Страница 108: ...5 Hardware operation 100 NEW KAVO 216508 rev 1 5 12 2 Panoramic imaging chain...
Страница 109: ...5 Hardware operation 216508 rev 1 NEW KAVO 101 5 12 3 Cephalometric imaging chain...
Страница 110: ...5 Hardware operation 102 NEW KAVO 216508 rev 1 5 12 4 Exposure control...
Страница 112: ...5 Hardware operation 104 NEW KAVO 216508 rev 1...
Страница 122: ...6 The s2terminal and service functions 114...
Страница 192: ...8 Maintenance procedure 184...