6 The s2terminal and service functions
110
-
off
state indicates that firmware upgrades have
been completed and the unit must be restarted for
the upgrades to be activated.
cancel
Skips the current initialization phase, i.e. initialization of
calibration data or initialization movements. This will allow
the unit to go from state init to idle without all initialization
being done.
Note!
This command can be used only for trouble shooting
purposes. The unit will not be fully operational if all initializa-
tions are not done!
patient
Changes the state of the unit from
idle
to
select
so that
you do not have to activate image capture to start an
examination.
service
Changes the state of the unit from
idle
or
error
to the
service state so that the unit service commands can be
accessed. (See section - Unit service commands)
reset
Changes the state of the unit from any state to the
init
state. This command reboots the firmware (not core) on
R3220 CPU.
clients
Show how many client applications are connected to the
unit, for example, the s2terminal and the dental imaging
software.
login
Shows the firmware and core versions installed in the unit
and the serial number of the unit.
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X ray System Service Manual ENGLISH 216508 rev 1 0 805 5127...
Страница 2: ......
Страница 4: ......
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 28: ...2 Unit description 20 2 6 Unit movements Panoramic unit movements R C J D...
Страница 31: ...2 Unit description 23...
Страница 32: ...2 Unit description 24...
Страница 37: ...3 Electrical description 29 Circuit boards Cephalometric unit...
Страница 54: ...46...
Страница 87: ...79...
Страница 88: ...80...
Страница 96: ...4 Firmware description 88...
Страница 101: ...5 Hardware operation 216508 rev 1 NEW KAVO 93 5 6 Z movement 2...
Страница 105: ...5 Hardware operation 216508 rev 1 NEW KAVO 97 5 10 Secondary collimator movement 2...
Страница 106: ...5 Hardware operation 98 NEW KAVO 216508 rev 1 5 11 Ceph sensor movement 2...
Страница 107: ...5 Hardware operation 216508 rev 1 NEW KAVO 99 5 12 Imaging chain 5 12 1 3D imaging chain...
Страница 108: ...5 Hardware operation 100 NEW KAVO 216508 rev 1 5 12 2 Panoramic imaging chain...
Страница 109: ...5 Hardware operation 216508 rev 1 NEW KAVO 101 5 12 3 Cephalometric imaging chain...
Страница 110: ...5 Hardware operation 102 NEW KAVO 216508 rev 1 5 12 4 Exposure control...
Страница 112: ...5 Hardware operation 104 NEW KAVO 216508 rev 1...
Страница 122: ...6 The s2terminal and service functions 114...
Страница 192: ...8 Maintenance procedure 184...