
7 Error codes and troubleshooting
132
7.4.20
Error 35 Movement of J (chin support)
motor failed
CAUSE A
The chin support is not moving correctly.
SOLUTION A
The movement not working correctly may be due to
a fault in the
• motor,
• motor cable,
• R3400 IO board.
Use service command “drivemotor j manual” to test
whether the motor is working properly. If the chin
support is moving the problem is probably in posi-
tion detection, see CAUSE B. Otherwise one of the
components listed above is faulty. Continue with in-
specting cables and connectors. Replace parts one
by one if needed.
CAUSE B
Position detection showing incorrect values.
SOLUTION B
Use service command “drivemotor j manual” and
the chin support up/down buttons to drive past the
positions where the opto switches are activated/de-
activated. Check for messages in the s2terminal in-
dicating the switches changing state. If no such
messages are seen, there is a problem with the
EG100 opto board or the cable between the opto
switch and the R3400 IO board. Check the parts
listed and replace if needed.
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X ray System Service Manual ENGLISH 216508 rev 1 0 805 5127...
Страница 2: ......
Страница 4: ......
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 28: ...2 Unit description 20 2 6 Unit movements Panoramic unit movements R C J D...
Страница 31: ...2 Unit description 23...
Страница 32: ...2 Unit description 24...
Страница 37: ...3 Electrical description 29 Circuit boards Cephalometric unit...
Страница 54: ...46...
Страница 87: ...79...
Страница 88: ...80...
Страница 96: ...4 Firmware description 88...
Страница 101: ...5 Hardware operation 216508 rev 1 NEW KAVO 93 5 6 Z movement 2...
Страница 105: ...5 Hardware operation 216508 rev 1 NEW KAVO 97 5 10 Secondary collimator movement 2...
Страница 106: ...5 Hardware operation 98 NEW KAVO 216508 rev 1 5 11 Ceph sensor movement 2...
Страница 107: ...5 Hardware operation 216508 rev 1 NEW KAVO 99 5 12 Imaging chain 5 12 1 3D imaging chain...
Страница 108: ...5 Hardware operation 100 NEW KAVO 216508 rev 1 5 12 2 Panoramic imaging chain...
Страница 109: ...5 Hardware operation 216508 rev 1 NEW KAVO 101 5 12 3 Cephalometric imaging chain...
Страница 110: ...5 Hardware operation 102 NEW KAVO 216508 rev 1 5 12 4 Exposure control...
Страница 112: ...5 Hardware operation 104 NEW KAVO 216508 rev 1...
Страница 122: ...6 The s2terminal and service functions 114...
Страница 192: ...8 Maintenance procedure 184...