DIAGNOSTICS 7-5
_______________________________________________________________________
Copyright © 2010 by Kaufman & Robinson, Inc., 1306 Blue Spruce Drive, Unit A, Fort Collins, CO 80524
Tel: 970-495-0187, Fax: 970-484-9350, Internet: www.ionsources.com
Help Code
or Error
Message
Description
Possible Causes And
Corrective Action
HLP 12
Initial Gas Failure. The
gas
fl
ow did not initialize
during the start up se-
quence.
Press and release the STAND-
BY/ENABLE button or remove
the remote enable signal to
clear the code. The unit remains
in STANDBY until STANDBY/
ENABLE is pressed again or the
remote enable signal is reap-
plied.
Gas bottle is empty and needs
to be replaced.
Make sure that any valves in the
system are open.
MFC is faulty or MFC cable is
faulty or disconnected. Repair
or replace MFC or cable. Make
sure MFC cable is connected.
HLP 13
Master Communication
Error.
Switch off the power until the
display is dark and then switch
the power back on to reset.