GENERAL DESCRIPTION AND SPECIFICATIONS 2-1
_______________________________________________________________________
Copyright © 2010 by Kaufman & Robinson, Inc., 1306 Blue Spruce Drive, Unit A, Fort Collins, CO 80524
Tel: 970-495-0187, Fax: 970-484-9350, Internet: www.ionsources.com
2 GENERAL DESCRIPTION AND SPECIFICATIONS
The auto controller package combines modular power supplies and automatic con-
trol for operating an End-Hall ion source. The auto controller is designed for use in
a standard 19-inch (48 cm) rack mount cabinet. The Auto Controller, Filament Pow-
er Supply, Discharge Power Supply and Mass Flow Controller(s) are included in
the standard package along with the associated interconnection and power cables.
The operating speci
fi
cations, weight and dimensions of each controller are outlined
in the following table.
This section gives a general overview of the Auto Controller package with more
detailed information in the later sections of this manual. Additional information on
each controller is given in the Appendix/Manual for each.
Table 2-1 Controller Speci
fi
cations
Nominal
Input voltage
Maximum
Input
Current
Maximum
Output
Weight
Height x
Depth
Auto
Controller
115 V, 1
Φ
, 50-60 Hz
208-230 V, 1
Φ
,
50-60 Hz
0.35 A
0.75 A
+/- 15 V,
500 mA DC
3.0 kg
(8 lbs)
4.4 x 53 cm
(1.75 x 21 in.)
Filament
Power
Supply
115 V, 1
Φ
, 50-60 Hz
208-230 V, 1
Φ
,
50-60 Hz
7.5 A
15 A
40 V,
25 A, AC
7.27 kg
(16 lbs)
4.5 x 53 cm
(1.75x 21in.)
Discharge
Power
Supply
115 V, 1
Φ
, 50-60 Hz
208-230 V, 1
Φ
,
50-60 Hz
15 A*
**
7.27 kg
(16 lbs)
4.5 x 53 cm
(1.75x 21in.)
* The discharge power supply output will be limited to 1500 watts when operation
with a nominal 115 volt ac input.
** The maximum output depends on the speci
fi
c power supply ordered.