3-10
12.06 / V1.0
AUTOCON
®
II 400
MODEL 205352 20-1
Error code
Description of error
Solution
Priority
C-75
Defective instrument interface
Check instrument and instrument cable
and reconnect
Low
C-7D
CRC error in program memory
Arrange for Black Box to be tested and
replaced if necessary
Low
C-7E
EEPROM not ready to read
Arrange for Black Box to be tested and
replaced if necessary
Low
C-7F
Multitasking error
Arrange for Black Box to be tested and
replaced if necessary
Low
C-81
Incorrect protocol length
Arrange for Black Box to be tested and
replaced if necessary
Low
C-82
SIO protocol error
Arrange for Black Box to be tested and
replaced if necessary
Low
C-83
No proper respond from instrument
recognized
Arrange for Black Box to be tested and
replaced if necessary
Low
C-84
Short circuit between AE and NE
Arrange for Black Box to be tested and
replaced if necessary
Low
C-91
Erroneous redundant current
measurement
Arrange for Black Box to be tested and
replaced if necessary
Low
C-92
No HF current with redundant
measurement
Arrange for Black Box to be tested and
replaced if necessary
Low
C-A0
Defective temperature sensor
Arrange for Black Box to be tested and
replaced if necessary
Low
C-A1
Erroneous EEPROM calibration
Arrange for Black Box to be tested and
replaced if necessary
Low
C-F0
Erroneous trial activation
Arrange for Black Box to be tested and
replaced if necessary
Low
I-01
Interrupt error
Arrange for Black Box to be tested and
replaced if necessary
Low
I-02
CRC error after reset
Arrange for Black Box to be tested and
replaced if necessary
Low
I-03
+5 V too low
Arrange for Black Box to be tested and
replaced if necessary
Low
I-04
+5 V too high
Arrange for Black Box to be tested and
replaced if necessary
Low
I-05
Undervoltage of ±12 V
Arrange for Black Box to be tested and
replaced if necessary
Low
I-06
Overvoltage of ±12 V
Arrange for Black Box to be tested and
replaced if necessary
Low
I-07
Current for activation recognition is too
low
Arrange for Black Box to be tested and
replaced if necessary
Low
I-08
Current for activation recognition is too
high
Arrange for Black Box to be tested and
replaced if necessary
Low
I-09
Current for instrument recognition is too
low
Arrange for Black Box to be tested and
replaced if necessary
Low
I-0A
Current for instrument recognition is too
high
Arrange for Black Box to be tested and
replaced if necessary
Low
I-0C
Stack error
Arrange for Black Box to be tested and
replaced if necessary
Low
I-0D
Error in state variable
Arrange for Black Box to be tested and
replaced if necessary
Low
I-0E
CRC error during operation
Arrange for Black Box to be tested and
replaced if necessary
Low
I-0F
Coag. voltage too low
Arrange for Black Box to be tested and
replaced if necessary
Low
I-70
Relay error
Arrange for Black Box to be tested and
replaced if necessary
Low
I-71
Relay error
Arrange for Black Box to be tested and
replaced if necessary
Low
I-72
Relay error
Arrange for Black Box to be tested and
replaced if necessary
Low
Содержание AUTOCON II 400
Страница 8: ......
Страница 9: ...AUTOCON II 400 AUTOCON II 400 AUTOCON II 400 20 5352 20 11x 20 5352 20 11x 20 5352 20 11x...
Страница 10: ......
Страница 13: ...III Ger teabbildungen Photographs of the equipment Im genes del equipo...
Страница 14: ...IV W R Q Ger teabbildungen Photographs of the equipment Im genes del equipo I O Y U T E...
Страница 146: ......
Страница 152: ......
Страница 156: ......
Страница 158: ......
Страница 172: ......
Страница 186: ...Figure 4 11 Figure 4 12 Figure 4 13 4 14 12 06 V1 0 AUTOCON II 400 MODEL 205352 20 1...
Страница 192: ......
Страница 198: ......
Страница 200: ......
Страница 202: ......
Страница 203: ......