AUTOCON
®
II 400
MODEL 205352 20-1
Work to be carried out
Remarks
쐽
Electric safety measurements
(IEC EN 60601-1)
– Protective ground resistance:
0.20
(with power cord)
– Earth leakage current:
– N.C.
0.50 mA
– S.F.C.
1.00 mA
– Housing leakage current: – N.C.
0.10 mA
– S.F.C.
0.50 mA
– Patient leakage current:
– N.C.
0.01 mA
– S.F.C.
0.05 mA
– Patient leakage current:
– S.F.C.
0.05 mA
(line voltage on the applied part)
N.C. = Normal condition
S.F.C. = Single fault condition
Caution:
Always unplug the device from the power supply before carrying out any
repairs on it!
After servicing, a safety test according to VDE 0751 or the leakage current and
protective ground resistance measurements according to IEC 60601-1,
UL-2601.1, CAN/CSA C22.2 NO 601.1 are to be carried out.