80
AUTOCON
®
II 400 Error Messages
Technische Beschreibung
Fehlersuche
Technical description
Troubleshooting
Error code
S
ymptom
Trouble
s
hooting
Priority
Alarm condition
A-0B:
NE contact: Resistance too low
Check neutral electrode and reposition if necessary
Medium
physiological
A-0C:
NE contact: Resistance too high
Check neutral electrode and reposition if necessary
Medium
physiological
A-10:
End activation (AutoStop reached)
End activation of HF energy
Low
other
A-11:
No activation signal
Have unit serviced.
High
technical
A-12:
Activation during power up not permitted
Check footswitch, handle and cable
Medium
other
A-17:
Simultaneous activation not permitted
Activate either via footswitch or handle
Medium
other
A-19:
Line voltage fault present
Check power cord, power outlet and power supply
Low
technical
A-1C:
Max. activation time exceeded
Cancel activation
Low
other
A-1E:
Touch screen pressed during power up
Touch screen must not be pressed during power-up
Low
technical
A-1F:
NE symmetry critical (cancel activation)
Check neutral electrode and reposition if necessary
High
technical
A-22:
AutoStop reached
Break contact with tissue
Low
other
A-85:
Dual-pedal footswitch disconnected
Reconnect dual-pedal footswitch and check cable
Low
technical
A-89:
Single-pedal footswitch disconnected
Reconnect single-pedal footswitch and check cable
Low
technical
A-94:
Triple-pedal footswitch disconnected
Reconnect triple-pedal footswitch and check cable
Low
technical
A-95:
Instrument detected at MF socket
New MF instrument at socket
Low
technical
A-9F:
MF instrument disconnected
MF instrument removed from socket
Low
technical
A-A3:
Footswitch not assigned to a socket
Assign footswitch to a socket (p. 56)
Low
other
A-A4:
2 footswitches of same type connected
Either connect one footswitch or two of a different type
Low
other
A-A9:
First select procedure before activation
Select the required procedure (p. 52)
Low
other
A-AA
No mode assigned
Assign a mode to the procedure (p. 53)
Low
other
A-AB
No/Unsuitable MF instrument detected
Attempt to activate the socket where there was no or a defective MF
instrument detected.
Medium
technical
A-AC
Autostart on tissue contact not possible
Release tissue contact, separate instrument tips, if necessary, assign
Autostart
Medium
other
A-B0
NE symmetry not optimal
Check neutral electrode and reposition if necessary
Low
technical
A-B7
Autostart only permitted up to 50W output
The max. output with Autostart should only be 50 W
Low
technical
A-BB
SC due
Arrange date for safety check
Medium
technical
A-C0
Activation type for MF-instrument missing
Specify activation type
Low
other
A-FE
Line voltage fault present
Check power cord, power outlet and power supply
Low
technical
C-84
Short-circuit between AE and NE
No impedance modification of tissue during coagulation
Low
technical
all others
Inform Service
Low
Descripción técnica
Localización de errores
Содержание AUTOCON II 400
Страница 8: ......
Страница 9: ...AUTOCON II 400 AUTOCON II 400 AUTOCON II 400 20 5352 20 11x 20 5352 20 11x 20 5352 20 11x...
Страница 10: ......
Страница 13: ...III Ger teabbildungen Photographs of the equipment Im genes del equipo...
Страница 14: ...IV W R Q Ger teabbildungen Photographs of the equipment Im genes del equipo I O Y U T E...
Страница 146: ......
Страница 152: ......
Страница 156: ......
Страница 158: ......
Страница 172: ......
Страница 186: ...Figure 4 11 Figure 4 12 Figure 4 13 4 14 12 06 V1 0 AUTOCON II 400 MODEL 205352 20 1...
Страница 192: ......
Страница 198: ......
Страница 200: ......
Страница 202: ......
Страница 203: ......