Error code
Description of error
Solution
Priority
A-95
Instrument recognized at MF socket
New MF instrument at socket
Low
A-98
CRC error at RAM self test
Medium
A-9A
Incompatible MF instrument
Low
A-9F
MF instrument disconnected
MF instrument disconnected from
socket
Low
A-A1
Missing status message from HF
module
High
A-A3
Footswitch is not assigned to a socket
Assign footswitch to a socket
Low
A-A4
Two identical footswitches connected
Either connect one footswitch or two of
different types
Low
A-A9
Select procedure prior to activation
Select any procedure
Low
A-AA
No mode assigned
Assign a mode to the procedure
Low
A-AB
No/Wrong MF instrument detected
Attempt activation at a socket at which
either no instrument or an invalid
instrument has been detected
Medium
A-AC
Autostart cannot be assigned at tissue
contact
Break tissue contact, disconnect
sections of the instrument as required,
assign autostart
Medium
A-AF
NE monitor ceases to respond
High
A-B0
Bad NE symmetry
Check neutral electrode and reposition
if necessary
Low
A-B1
NE current density too high
Medium
A-B7
Power is limited to 50 W with autostart
The max. output with autostart should
only be 50 W
Low
A-BB
Next safety check date overdue
Agree date for safety check
Medium
A-C0
No activation type for MF socket
selected
Specify activation type
Low
A-C1
Activation not allowed
Low
A-C2
Defective front panel fan
Low
A-C3
Select new mode for MF instrument
A-FB
Socket configuration has been changed
Low
A-FE
Power fail signal detected
Check power cord, power outlet and
power supply
Low
A-FF
Invalid password
Low
C-01
Missing ESI message from front panel
Arrange for Black Box to be tested and
replaced if necessary
Low
C-04
Internal status error
Arrange for Black Box to be tested and
replaced if necessary
Low
C-05
Wrong ECB message length
Arrange for Black Box to be tested and
replaced if necessary
Low
C-06
No parameter received
Arrange for Black Box to be tested and
replaced if necessary
Low
C-0D
Wrong ECB message length
Arrange for Black Box to be tested and
replaced if necessary
Low
C-21
Erroneous power dosage
Arrange for Black Box to be tested and
replaced if necessary
Low
C-26
Erroneous power measurement
Arrange for Black Box to be tested and
replaced if necessary
Low
C-30
HF voltage too high (redundant)
Arrange for Black Box to be tested and
replaced if necessary
Low
C-31
HF voltage too high (OFF)
Arrange for Black Box to be tested and
replaced if necessary
Low
C-32
HF voltage too low (OFF)
Arrange for Black Box to be tested and
replaced if necessary
Low
C-33
HF voltage too high (ON)
Arrange for Black Box to be tested and
replaced if necessary
Low
3-8
12.06 / V1.0
AUTOCON
®
II 400
MODEL 205352 20-1
Содержание AUTOCON II 400
Страница 8: ......
Страница 9: ...AUTOCON II 400 AUTOCON II 400 AUTOCON II 400 20 5352 20 11x 20 5352 20 11x 20 5352 20 11x...
Страница 10: ......
Страница 13: ...III Ger teabbildungen Photographs of the equipment Im genes del equipo...
Страница 14: ...IV W R Q Ger teabbildungen Photographs of the equipment Im genes del equipo I O Y U T E...
Страница 146: ......
Страница 152: ......
Страница 156: ......
Страница 158: ......
Страница 172: ......
Страница 186: ...Figure 4 11 Figure 4 12 Figure 4 13 4 14 12 06 V1 0 AUTOCON II 400 MODEL 205352 20 1...
Страница 192: ......
Страница 198: ......
Страница 200: ......
Страница 202: ......
Страница 203: ......